202B1263075
Problem
Cause
Remedy
Loss of image during
an examination or
treatment.
1) The imaging section is damaged.
2) The scope connection is incomplete.
3) The system is malfunctioning possibly due
to static charge.
4) The video signal cable has shorted out or
broken.
1) If the problem occurred during
treatment
[Note 1]
, stop treatment and
remove the endotherapy device from
the endoscope. Then, reset
[Note]
the video
processor and the light source. If the
problem occurred during an examination,
reset
[Note]
the video processor and the light
source. If an appropriate image does not
appear even after resetting, turn them off,
straighten the bending section to unlock,
release the angle knobs, withdraw the
endoscope slowly from the patient, and
then contact your local FUJIFILM dealer.
2) Reconnect the endoscope.
“3.6 Connecting Endoscope”
3) 4)
If the problem occurred during
treatment
[Note 1]
, stop treatment and remove
the endotherapy device from the endoscope.
Then, reset
[Note]
the video processor and
the light source. If the problem occurred
during an examination, reset
[Note]
the
video processor and the light source. If an
appropriate image does not appear even
after resetting, turn them off, straighten the
bending section to unlock, release the angle
knobs, withdraw the endoscope slowly
from the patient, and then contact your local
FUJIFILM dealer.
Live image is not
displayed after image
freezing is cancelled.
The system is malfunctioning possibly due to
static charge.
If the problem occurred during treatment
[Note 1]
,
stop treatment and remove the endotherapy
device from the endoscope. Then, reset
[Note]
the video processor and the light source. If
the problem occurred during an examination,
reset
[Note]
the video processor and the light
source. If an appropriate image does not
appear even after resetting, turn them off,
straighten the bending section to unlock,
release the angle knobs, withdraw the
endoscope slowly from the patient, and then
contact your local FUJIFILM dealer.
[Note] Reset: Turn off the video processor and the light source, and wait for at least 5 seconds. Turn on the video
processor and the light source again, and then light the lamp by pressing the Lamp button.
[Note 1] During treatment with a snare, withdraw the snare after loosening it enough.