3.7.4 Inspecting the Balloon Air Feed Inlet
Supply air from the balloon air feed inlet with a syringe
and make sure that the air exits from the balloon air feed
outlet.
3.7.5 Inspecting the Instrument Channel
Insert an endotherapy device from the instrument channel
inlet and check that the tip of endotherapy device comes
smoothly out of the instrument channel outlet in the distal
end of the endoscope.