Figure 10-2. Tunnel Diode Pulser Connections
The Leveled Sine Wave Function
The Leveled Sine Wave (LEVSINE) function uses a leveled sine wave, whose amplitude
remains relatively constant over a range of frequencies, to check the oscilloscope’s
bandwidth. When you check your oscilloscope, you change the wave’s frequency until
the amplitude displayed on the oscilloscope drops 30 %, which is the amplitude that
corresponds to the -3 dB point. Default values are 30 mV p-p, 50 kHz.
To access the LEVSINE menu, press the softkey under MODE until “levsine” appears.
gjh056.eps
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press
P
to return directly to the SCOPE menu.
Each option in the LEVSINE menu is described below:
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...