5522A
Operators Manual
1-12
AC Voltage (Sine Wave)
Absolute Uncertainty,
tcal
±
5
°
C
±
(ppm of
μ
V)
Range Frequency
90 days
1 year
Resolution
Max
Burden
Max Distortion and
Noise
10 Hz to 5 MHz
Bandwidth
±
(% floor)
Normal Output
10 Hz to 45 Hz
600 + 6
800 + 6
0.15 + 90
μ
V
45 Hz to 10 kHz
120 + 6
150 + 6
0.035 + 90
μ
V
10 kHz to 20 kHz
160 + 6
200 + 6
0.06 + 90
μ
V
20 kHz to 50 kHz
800 + 6
1000 + 6
0.15 + 90
μ
V
50 kHz to 100 kHz
3000 + 12
3500 + 12
0.25 + 90
μ
V
1.0 mV to
32.999 mV
100 kHz to 500 kHz
6000 + 50
8000 + 50
1
μ
V 65
Ω
0.3 + 90
μ
V
[1]
10 Hz to 45 Hz
250 + 8
300 + 8
0.15 + 90
μ
V
45 Hz to 10 kHz
140 + 8
145 + 8
0.035 + 90
μ
V
10 kHz to 20 kHz
150 + 8
160 + 8
0.06 + 90
μ
V
20 kHz to 50 kHz
300 + 8
350 + 8
0.15 + 90
μ
V
50 kHz to 100 kHz
600 + 32
800 + 32
0.20 + 90
μ
V
33 mV to
329.999 mV
100 kHz to 500 kHz
1600 + 70
2000 + 70
1
μ
V 65
Ω
0.20 + 90
μ
V
[1]
10 Hz to 45 Hz
250 + 50
300 + 50
0.15 + 200
μ
V
45 Hz to 10 kHz
140 + 60
150 + 60
0.035 + 200
μ
V
10 kHz to 20 kHz
160 + 60
190 + 60
0.06 + 200
μ
V
20 kHz to 50 kHz
250 + 50
300 + 50
0.15 + 200
μ
V
50 kHz to 100 kHz
550 + 125
700 + 125
0.20 + 200
μ
V
0.33 V to
3.29999 V
100 kHz to 500 kHz
2000 + 600
2400 + 600
10
μ
V
10 mA
0.20 + 200
μ
V
[1]
10 Hz to 45 Hz
250 + 650
300 + 650
0.15 + 2 mV
45 Hz to 10 kHz
125 + 600
150 + 600
0.035 + 2 mV
10 kHz to 20 kHz
220 + 600
240 + 600
0.08 + 2 mV
20 kHz to 50 kHz
300 + 600
350 + 600
0.2 + 2 mV
3.3 V to
32.9999 V
50 kHz to 100 kHz
750 + 1600
900 + 1600
100
μ
V
10 mA
0.5 + 2 mV
45 Hz to 1 kHz
150 + 2000
190 + 2000
0.15 + 10 mV
1 kHz to 10 kHz
160 + 6000
200 + 6000
0.05 +10 mV
10 kHz to 20 kHz
220 + 6000
250 + 6000
0.6 + 10 mV
20 kHz to 50 kHz
240 + 6000
300 + 6000
0.8 + 10 mV
33 V to
329.999 V
50 kHz to 100 kHz
1600 + 50000 2000 + 50000
1 mV
5 mA,
except
20 mA for
45 Hz to
65 Hz
1.0 + 10 mV
45 Hz to 1 kHz
250 + 10000
300 + 10000
0.15 + 30 mV
1 kHz to 5 kHz
200 + 10000
250 + 10000
0.07 + 30 mV
330 V to
1020 V
5 kHz to 10 kHz
250 + 10000
300 + 10000
10 mV
2 mA,
except 6 mA
for 45 Hz to
65 Hz
0.07 + 30 mV
[1]
Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9 % of floor as shown.
Note
Remote sensing is not provided. Output resistance is <5 m
Ω
for outputs
≥
0.33 V. The AUX output resistance is <1
Ω
. The maximum
load capacitance is 500 pF, subject to the maximum burden current limits
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...