Remote Operations
How to Set up the RS-232 Host Port for Remote Control
5
5-9
8.
Verify that the calibrator is now in remote control.
9.
Select the Local command from the Control menu, then click OK in the Parameter
Input Window. Observe the Calibrator Control Display changes back to the reset
condition (below).
nn323f.eps
10.
From the ud0: prompt, type <q> and then press the ENTER (or RETURN) key.
How to Set up the RS-232 Host Port for Remote Control
The Calibrator is fully programmable over an RS-232 link with a PC the rear panel
SERIAL 1 FROM HOST port (Figure 5-2). You can enter individual commands from a
terminal, write your own programs using, for example, a Windows-based language such
as Visual Basic, or run optional Windows-based Fluke software such as 5500/CAL or
MET/CAL.
The RS-232 cable length for the port should not exceed 15 meters (50 feet), although
longer cable lengths are permitted if the load capacitance measured at a connection point
(including signal terminator) does not exceed 2500 pF.
RS-232 Host Port Setup Procedure
Complete the following procedure to set up the SERIAL 1 FROM HOST port. The
RS-232 parameters you select here must match the parameters set for the PC COM port.
The factory defaults (shown on the display below) are 9600 baud, 8 data bits, 1 stop bit,
and no parity. Other parameters include flow control, EOL (end-of-line) character, and
EOF (end-of-file) characters.
1.
Turn the Calibrator power on. You may operate the Calibrator during warmup, but
specifications are not guaranteed until warmup is complete.
2.
Press
S
on the Calibrator front panel.
3.
Negotiate the softkey selections shown below to select the serial port for remote
operation, then continue to Step 4.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...