5522A
Operators Manual
6-44
Example:
TEMP_STD ITS_90
See the temperature standard to ITS-90.
TEMP_STD?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Temperature Degree Standard command) Returns the temperature standard ipts-68
(1968 International Provisional Temperature Standard) or its-90 (1990 International
Temperature Standard).
Parameters:
IPTS_68
ITS_90
Example:
TEMP_STD?
returns
ITS_90
Returns
ITS_90
when the temperature degree standard is the 1990 International
Temperature Standard.
*TRG
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Trigger Thermocouple Measurement command) Triggers a thermocouple temperature
measurement and return the value of the measurement. Also changes the operating mode
to thermocouple measurement if this is not already the operating mode. (This command is
equivalent to sending
TC_MEAS;*WAI;VAL?
)
Responses: <measurement
value>,
CEL
(value is in Celsius)
<measurement
value>,
FAR
(value is in Fahrenheit)
0.00E+00,OVER
(value is over or under capability)
0.00E+00,OPENTC
(open
thermocouple)
0.00E+00,NONE
(wrong mode or no measurement)
Example:
*TRG
returns
+2.500E+01,CEL
Trigger a thermocouple measurement and return 25.00 Celsius when the thermocouple
temperature measurement is 25
°
C.
TSENS_TYPE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Temperature Sensor Type command) Sets the temperature sensor type to thermocouple
(TC) or Resistance Temperature Detector (RTD) for temperature measurements. The
Calibrator simulates the RTD temperature as a resistance output on the NORMAL
terminals, and simulates the thermocouple temperature as a dc voltage output on the TC
terminals. If the temperature sensor type is changed, the temperature output is reset to 0
degrees C. Once set, the Calibrator retains the temperature sensor type until power off or
reset.
Parameters:
TC
(Thermocouple)
RTD
(Resistance
Temperature
Detector)
Example:
TSENS_TYPE RTD
Set the temperature sensor type to an RTD.
TSENS_TYPE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Temperature Sensor Type query) Returns the temperature sensor type thermocouple
(TC) or Resistance Temperature Detector (RTD) for temperature measurements.
Responses:
TC
(Thermocouple)
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...