Remote Commands
Command Summary by Function
6
6-9
Table 6-9. Setup and Utility Commands (cont.)
Command Description
REFCLOCK_D?
Returns the power-up and reset default for the reference clock source (internal or
through the 10 MHz IN BNC connector).
REFPHASE_D
If two Calibrators are synchronized using 10 MHz IN/OUT, sets the power-up and reset
default phase difference between the NORMAL terminals on the slave Calibrator and the
NORMAL terminals of the master Calibrator.
REFPHASE_D?
If two Calibrators are synchronized using 10 MHz IN/OUT, returns the power-up and
reset default phase difference between the NORMAL terminals on the slave Calibrator
and the NORMAL terminals of the master Calibrator.
RTD_TYPE_D
Set the default Resistance Temperature Detector (RTD) sensor type.
RTD_TYPE_D? Returns the default Resistance Temperature Detector (RTD) sensor type.
SP_SET
Sets the HOST serial port communication parameters and saves them in nonvolatile
memory.
SP_SET?
Returns the HOST serial port communication parameters contained in nonvolatile
memory.
TC_TYPE_D
Sets the power-up and reset default thermocouple type.
TC_TYPE_D?
Returns the power-up and reset default thermocouple type.
TEMP_STD
Sets the temperature degree standard, ipts-68 or its-90.
TEMP_STD?
Returns the temperature degree standard, ipts-68 or its-90.
TLIMIT_D
Sets the power-up and reset default for the time limit for –SC600 OVERLD mode to stay
in operate.
TLIMIT_D?
Returns the power-up and reset default for the time limit for –SC600 OVERLD mode to
stay in operate.
UNCERT?
Retums specified uncertainties for the present output. If there are no specifications for an
output, returns zero.
Table 6-10. Status Commands
Command Description
ERR?
Returns the first error code with an explanation contained in the Calibrator error queue,
then removes that error code from the queue.
EXPLAIN?
Explains an error code. This command returns a string that explains the error code
furnished as the parameter.
FAULT?
Returns the first error code contained in the Calibrator error queue, then removes that
error from the queue.
FUNC?
Returns the present output, measurement, or calibration function.
ISCE
Loads two bytes into both the Instrument Status 1 to 0 Change Enable register and the
Instrument Status 0 to 1 Change Enable register.
ISCE?
Returns the OR of the contents of the Instrument Status 1 to 0 Change Enable register
and the Instrument Status 0 to 1 Change Enable register.
ISCE0
Loads two bytes into the Instrument Status 1 to 0 Change Enable register.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...