Remote Operations
How to Set up the RS-232 UUT Port for Remote Control
5
5-19
UUT
5522A Cali
b
rator
Controller
SERIAL 2
TO UUT
Port
RS-232
Port
Modem Ca
b
le
IEEE-4
88
Ca
b
le
gjh046.eps
Figure 5-6. Testing the RS-232 UUT Port via IEEE-488 Port
Complete the following procedure to test RS-232 UUT port operation via the IEEE-488
port using the Win32 Interactive Control utility.
1.
Complete the “IEEE-488 Port Setup Procedure” earlier in this chapter to set up the
Calibrator for GPIB operation.
2.
Complete “Testing the IEEE-488 Port” to prepare the Calibrator IEEE-488 port for
testing. Before the final step, return to this procedure and continue to Step 3 below.
3.
Go to Start then to the Programs menu.
4.
Select "NI-488.2M software for... (your operating system)".
5.
From the NI488.2M software menu, select "Win32 interactive control".
6.
A DOS window opens with a prompt as shown here:
7.
At the prompt, type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
8.
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...