How to Test Video Triggers
gjh062.eps
The video mode generates video signals in various formats. The mode is used to test the
video trigger capability of an oscilloscope. You can press the MODE softkey to cycle
through the functions in the order shown, or you can press
P
to return directly to the
OTHER modes menu.
Each option in the VIDEO menu is described below:
•
Output @ SCOPE terminal (50
Ω
)
Indicates the location of the signal output. If the
signal does not appear on the oscilloscope, press
O
. To disconnect the signal,
press
Y
.
•
LINE MK
Allows you to select the marker line number. For ntsc and pal-m
formats, you can also select field (“odd” or “even”). For pal and secam formats, the
field (“ODD” or “EVEN”) is selected automatically based on marker line number.
•
FORMAT
Scrolls through the available formats. You can select ntsc, pal, pal-m,
and secam.
•
MODE
Indicates the calibrator is in VIDEO mode. Use the softkey to change modes
and open menus for other oscilloscope calibration modes.
Default video settings are + 100 %, format = NTSC, and videomark = 10.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...