5522A
Operators Manual
1-18
Temperature Calibration (RTD)
Absolute Uncertainty
tcal
±
5
°
C
±
°
C
[2]
Absolute Uncertainty
tcal
±
5
°
C
±
°
C
[2]
RTD Type
Range
°
C
[1]
90 days
1 year
RTD Type
Range
°
C
[1]
90 days
1 year
-200 to -80
0.04
0.05
-200 to -80
0.03
0.04
-80 to 0
0.05
0.05
-80 to 0
0.04
0.05
0 to 100
0.07
0.07
0 to 100
0.05
0.05
100 to 300
0.08
0.09
100 to 260
0.06
0.06
300 to 400
0.09
0.10
260 to 300
0.07
0.08
400 to 630
0.10
0.12
300 to 400
0.07
0.08
Pt 385,
100
Ω
630 to 800
0.21
0.23
400 to 600
0.08
0.09
-200 to -80
0.04
0.05
Pt 385,
500
Ω
600 to 630
0.09
0.11
-80 to 0
0.05
0.05
-200 to -80
0.03
0.03
0 to 100
0.07
0.07
-80 to 0
0.03
0.03
100 to 300
0.08
0.09
0 to 100
0.03
0.04
300 to 400
0.09
0.10
100 to 260
0.04
0.05
Pt 3926,
100
Ω
400 to 630
0.10
0.12
260 to 300
0.05
0.06
-200 to -190
0.25
0.25
300 to 400
0.05
0.07
-190 to -80
0.04
0.04
400 to 600
0.06
0.07
-80 to 0
0.05
0.05
Pt 385,
1000
Ω
600 to 630
0.22
0.23
0 to 100
0.06
0.06
-80 to 0
0.06
0.08
100 to 260
0.06
0.07
0 to 100
0.07
0.08
260 to 300
0.07
0.08
PtNi 385,
120
Ω
(Ni120)
100 to 260
0.13
0.14
300 to 400
0.08
0.09
400 to 600
0.08
0.10
Cu 427,
10
Ω
[3]
-100 to 260
0.3
0.3
Pt 3916,
100
Ω
600 to 630
0.21
0.23
-200 to -80
0.03
0.04
-80 to 0
0.03
0.04
0 to 100
0.04
0.04
100 to 260
0.04
0.05
260 to 300
0.11
0.12
300 to 400
0.12
0.13
400 to 600
0.12
0.14
Pt 385,
200
Ω
600 to 630
0.14
0.16
[1]
Resolution is 0.003
°
C
[2]
Applies for COMP OFF (to the 5522A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3]
Based on MINCO Application Aid No. 18
DC Power Specification Summary
Current Range
0.33 to
329.99 mA
0.33 to
2.9999 A
3 to
20.5 A
Voltage
Range
Absolute Uncertainty, tcal
±
5
°
C,
±
(% of watts output)
[1]
90 days
33 mV to 1020 V
0.021
0.019
[2]
0.06
[2]
1 year
33 mV to 1020 V
0.023
0.022
[2]
0.07
[2]
[1]
To determine dc power uncertainty with more precision, see the individual “AC Voltage Specifications,” “AC Current
Specifications,” and “Calculating Power Uncertainty.”
[2]
Add 0.02 % unless a settling time of 30 seconds is allowed for output currents >10 A or for currents on the highest two current
ranges within 30 seconds of an output current >10 A.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...