SC600 Oscilloscope Calibration Option
Remote Commands and Queries
9
9-27
Table 9-1. SCOPE Command Parameters (cont.)
Parameter Description/Example
EDGE
Oscilloscope EDGE mode. Programs 25 mV peak-to-peak, 1 MHz, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns EDGE.
Example:
SCOPE EDGE; OUT 0.5 V, 5 kHz
(Edge, 0.5 V peak-to-peak, 5 kHz.)
LEVSINE
Oscilloscope LEVSINE mode. Programs 30 mV peak-to-peak, 50 kHz, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns LEVSINE.
Example:
SCOPE LEVSINE; OUT 1 V, 50 kHz
(Leveled sine wave, 1 V peak-to-peak, 50 kHz.)
MARKER
Oscilloscope MARKER mode. Programs the period to 1 ms, output at the SCOPE
BNC, standby if from OFF or previously in standby. FUNC? returns MARKER.
Example:
SCOPE MARKER; OUT 2 MS
(Marker, period of 2 ms.)
WAVEGEN Oscilloscope
WAVEGEN
mode. Programs 20 mV peak-to-peak, square wave, 1 kHz,
no offset, output impedance 1 M
Ω
, standby if from OFF or previously in standby.
FUNC? returns WAVEGEN.
Example:
SCOPE WAVEGEN; OUT 1 V, 1 kHz
(Wave Generator, 1 V peak-to-peak, 1 kHz.)
VIDEO
Oscilloscope VIDEO mode. Programs 100% output (1 V p-p), line marker 10, format
NTSC. FUNC? returns VIDEO.
Examples:
SCOPE VIDEO; OUT 90
(Video, 90% output)
SCOPE VIDEO; OUT -70
(Video, -70% output, inverse video)
PULSE
Oscilloscope PULSE mode. Programs 100 ns pulse width, 1.000
μ
s period, 2.5 V
range. FUNC? returns PULSE.
Example:
SCOPE PULSE; OUT 50 ns, 500 ns; RANGE TP8DB
(Pulse, 50 ns pulse width, 500 ns period, 1.5 V range)
MEASZ Oscilloscope
Impedance/Capacitance
measurement (MEAS Z) mode. Programs 50
Ω
range. FUNC? returns MEASZ.
Example:
SCOPE MEASZ; RANGE TZCAP
(MEAS Z mode, capacitance range)
OVERLD
Oscilloscope Overload mode. Programs 5 V dc range. FUNC? returns OVERLD.
Example:
SCOPE
OVERLD; OUT 7 V; RANGE TOLAC
(Overload, 7 V output, ac range)
SCOPE
(IEEE-488, RS-232, Sequential)
Programs the 5500A-SC600 oscilloscope calibration option hardware, if installed. The
instrument settings are determined by this command’s parameter. Once in SCOPE mode,
use the OUT command to program new output.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...