SC600 Oscilloscope Calibration Option
General Specifications
9
9-5
Leveled Sine Wave Specifications
Frequency Range
Leveled Sine Wave
Characteristics into 50
Ω
50 kHz (reference) 50 kHz to 100 MHz
100 to 300 MHz
300 to 600 MHz
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p)
5 mV to 5.5 V
Resolution
<100 mV: 3 digits
≥
100 mV: 4 digits
Adjustment Range
continuously adjustable
1-Year Absolute
Uncertainty, tcal
±
5
°
C
±
(2 % of output
+ 300
μ
V)
±
(3.5 % of output
+ 300
μ
V)
±
(4 % of output
+ 300
μ
V)
±
(6 % of output
+ 300
μ
V)
Flatness (relative to
50 kHz)
not applicable
±
(1.5 % of output
+ 100
μ
V)
±
(2 % of output
+ 100
μ
V)
±
(4 % of output
+ 100
μ
V)
Short-Term Amplitude
Stability
≤
1 %
[1]
Frequency Characteristics
Resolution
1 kHz
10 kHz
1-Year Absolute
Uncertainty, tcal
±
5
°
C
±
2.5 ppm
[2]
Distortion Characteristics
2nd Harmonic
≤
-33 dBc
3rd and Higher Harmonics
≤
-38 dBc
[1]
Within 1 hour after reference amplitude setting, provided temperature varies no more than
±
5 °C.
[2] With REF CLK set to ext, the frequency uncertainty of the Leveled Sine Wave is the uncertainty of the external 10 MHz clock
±
0.3 Hz/gate time.
Time Marker Specifications
Time Maker into 50
Ω
5 s to 50 ms
20 ms to
100 ns
50 to 20 ns
10 ns
5 to 2 ns
1-Year Absolute
Uncertainty at Cardinal
Points, tcal
±
5
°
C
±
(25 + t *1000)
ppm
[1]
±
2.5 ppm
±
2.5 ppm
±
2.5 ppm
±
2.5 ppm
Wave Shape
spike or square
spike, square,
or 20 %-pulse
spike or square
square or sine
sine
Typical Output Level
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
[2]
>1 V p-p
Typical Jitter (rms)
<10 ppm
<1 ppm
<1 ppm
<1 ppm
<1 ppm
Sequence
5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
Adjustment Range
[3]
At least
±
10 % around each sequence value indicated above.
Amplitude Resolution
4 digits
[1]
t is the time in seconds.
[2]
Typical rise time of square wave and 20 %-pulse (20 % duty cycle pulse) is < 1.5 ns.
[3]
Time marker uncertainty is
±
50 ppm away from the cardinal points.
Wave Generator Specifications
Wave Generator Characteristics
Square Wave, Sine Wave, and Triangle Wave
into 50
Ω
or 1 M
Ω
Amplitude
Range
into 1 M
Ω
:
1.8 mV to 55 V p-p
into 50
Ω
:
1.8 mV to 2.5 V p-p
1-Year Absolute Uncertainty, tcal
±
5
°
C,
10 Hz to 10 kHz
±
(3 % of p-p 100 µV)
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical DC Offset Range
0 to
±
(
≥
40 % of p-p amplitude)
[1]
Frequency
Range
10 Hz to 100 kHz
Resolution
4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal
±
5
°
C
±
(25 ppm + 15 mHz)
[1]
The DC offset plus the wave signal must not exceed 30 V rms.
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...