5522A
Operators Manual
6-16
ERR?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Error query) Returns the first error code contained in the Calibrator error queue, then
removes that error code from the queue. Following the error code is an explanation of the
error code, similar to but sometimes containing more specific information than the
EXPLAIN?
command. The explanation sent in response to this query can contain
variables specific to a particular error event. See Appendix E for a list of error codes and
error messages.
A zero value is returned when the error queue is empty. To read the entire contents of the
error queue, repeat
ERR?
until the response
0,”No Error”
is returned. For terminal
users, the error queue Returns for
ERR?
is always
0,”No Error”
because error
messages are returned instead of queued.
Response:
<value>, (error code value)
<string>
(text string explaining the error)
Example:
ERR?
returns
0,”No Error”
Returns
0,”No Error”
when the error queue is empty.
ERR_REF
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Chooses the error reference for error calculations.
Parameter:
NOMINAL
Sets the reference to the nominal value
TRUVAL
Sets the reference to the output value
ERR_REF?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Returns the error reference for error calculations.
Response:
NOMINAL
The nominal value is used as the error reference
TRUVAL
The output value is used as the error reference
ERR_UNIT
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Error Unit Thresh Hold command) Chooses how UUT error is shown (this iS
nonvolatile).
Parameter:
GT1000
UUT error is displayed in % above 1000 ppm, ppm below
GT100
UUT error is displayed in % above 100 ppm, ppm below
GT10
UUT error is displayed in % above 10 ppm, ppm below
PPM
UUT error is displayed in ppm always
PCT
UUT error is displayed in % always
ERR_UNIT?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Error Unit Thresh Hold query) Returns presently selected values of ERR_UNIT.
Responses:
GT1000
UUT error is displayed in % above 1000 ppm, ppm below
GT100
UUT error is displayed in % above 100 ppm, ppm below
GT10
UUT error is displayed in % above 10 ppm, ppm below
PPM
UUT error is displayed in ppm always
PCT
UUT error is displayed in % always
*ESE
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Event Status Enable command) Loads a byte into the Event Status Enable (ESE)
register. (See “Event Status Enable Register (ESE)” in Chapter 5)
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...