5522A
Operators Manual
6-34
Returns 0 when the output is not being edited.
Example:
REFOUT?
returns
2.500000E-01
Returns .250 when the output is being edited and the reference is, for example, 250 mV.
REFPHASE
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Reference Phase command) If two Calibrators are synchronized using 10 MHz
IN/OUT, sets the phase difference between the primary channel on the Calibrator relative
to the sync pulse on the 10 MHz IN or OUT terminal. The primary channel is the
NORMAL, AUX, or 20A terminal for single outputs and the NORMAL terminal for ac
power and ac dual voltage outputs. The sync pulse on the 10 MHz IN or OUT terminal is
the phase reference. The set range is 0.00 to
±
180.00 degress, with + for a leading phase
difference and
−
for a lagging phase difference.
Parameter:
Phase with optional multiplier and DEG unit
Example:
REFPHASE 1.5 DEG
(1.5 degrees)
On either Calibrator, set the phase of the primary channel to lead the sync pulse by 1.5
degrees.
REFPHASE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Reference Phase query) If two Calibrators are synchronized using 10 MHz IN/OUT,
returns the phase difference between the primary channel on the Calibrator and the sync
pulse on the 10 MHz IN or OUT terminal.
Response:
(float) Phase in degrees
Example:
REFPHASE?
returns
1.50E+00
(1.5 degrees)
REFPHASE_D
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Reference Phase Default command) If two Calibrators are synchronized using 10 MHz
IN/OUT, sets the power-up and reset default phase difference between the primary
channel on the Calibrator relative to the sync pulse on the 10 MHz IN or OUT terminal.
The primary channel is the NORMAL, AUX, or 20A terminal for single outputs and the
NORMAL terminal for ac power and ac dual voltage outputs. The sync pulse on the
10 MHz IN or OUT terminal is the phase reference. The set range is 0.00 to
±
180.00 degrees, with + for a leading phase difference and
−
for a lagging phase
difference.
Parameter:
Phase with optional multiplier and DEG unit
Example
REFPHASE_D 1.5 DEG
(1.5 degrees)
On either Calibrator, set the power-up and reset default phase of the primary channel to
lead the sync pulse by 1.5 degrees.
REFPHASE_D?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Reference Phase Default query) If two Calibrators are synchronized using 10 MHz
IN/OUT, returns the power-up and reset default phase difference between the primary
channel on the Calibrator and the sync pulse on the 10 MHz IN or OUT terminal.
Response:
(Float) Phase in degrees
Example:
REFPHASE_D?
returns
1.50E+00
(1.5 degrees)
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...