Remote Commands
Commands
6
6-21
Load the error mode and increment the selected edit field by .00001 mV.
ISCE
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status Change Enable command) Loads two bytes into the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable Registers” in
Chapter 5 for more information.)
Parameter: <value>
(decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE 6272
Load decimal 6272 (binary
0001010001000000
) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT). This is equivalent to sending the commands
ISCE0 6272
and
ISCE1 6272
(see below).
ISCE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status Change Enable query) Returns the two bytes from the two 16-bit
ISCE mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable
Registers” in Chapter 5 for more information.)
Response: <value>
(decimal
equivalent of the 16 bits, 0 to 32767)
Example:
ISCE?
returns
6272
Returns decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCE0
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status 1 to 0 Change Enable command) Loads the two bytes into the 16-bit
ISCE0 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Parameter: <value>
(decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE0 6272
Load decimal 6272 (binary
0001010001000000
) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT).
ISCE0?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status 1 to 0 Change Enable query) Returns the two bytes from the 16-bit
ISCE0 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Response: <value>
(decimal
equivalent of the 16 bits, 0 to 32767)
Example:
ISCE0?
returns
6272
Returns decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCE1
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status 0 to 1 Change Enable command) Loads the two bytes into the 16-bit
ISCE1 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Parameter: <value>
(decimal equivalent of the 16 bits, 0 to 32767)
Содержание 522A/6
Страница 4: ......
Страница 14: ...5522A Operators Manual x...
Страница 18: ...5522A Operators Manual xiv...
Страница 22: ...5522A Operators Manual 1 2...
Страница 46: ...5522A Operators Manual 1 26...
Страница 48: ...5522A Operators Manual 2 2...
Страница 54: ...5522A Operators Manual 2 8...
Страница 56: ...5522A Operators Manual 3 2...
Страница 189: ...6 1 Chapter 6 Remote Commands Title Page Introduction 6 3 Command Summary by Function 6 3 Commands 6 10...
Страница 190: ...5522A Operators Manual 6 2...
Страница 240: ...5522A Operators Manual 7 2...
Страница 260: ...5522A Operators Manual 8 2...
Страница 314: ...5522A Operators Manual 9 52...
Страница 370: ...5522A Operators Manual 10 56...
Страница 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 414: ...5522A Operators Manual B 2...
Страница 416: ...5522A Operators Manual B 4...