High Performance Oscilloscope Calibrator
Model 9500B Test and Selftest
17
Selecting 'DISPLAY' transfers to the 'Display test.' screen. This invites a user to use the
front panel 'tab' (
Δ
) key to move the dark band down the screen, up to the top, and down
again, thus testing all the display elements on the screen:
erw187
The screen text will shift automatically so as not to obscure the band as it jumps back to
the top of the screen. If there are elements of the screen which do not show light in the
light background or dark in the dark band as it is advanced down the screen, then this
implies a failure. Rectification will require access to the internal circuitry, so no further
user action is recommended, except to report the result to your Fluke Calibration Service
Center.
The 'Enter' Key (
ξ
) returns to the Interface 'Select test' menu screen.
Memory Card Checks
Memory Card checks are initiated by pressing the MEMORY CARD key on the 'Select
test' menu screen:
erw188