High Performance Oscilloscope Calibrator
Error Messages
29
Device-Dependent Errors (DDE)
A Device-Dependent Error is generated if the device detects an internal operating fault
(eg. during self-test). The DDE bit (3) is set true in the Standard-defined Event Status
Byte, and the error code number is appended to the Error queue. The error description
appears on the display, remaining visible until the next key press or remote command.
Errors are reported by the mechanisms which deal with status reporting.
Device-Dependent Errors Reported only Locally on the Front Panel Screen
The error list for local operations, which are not reported to the remote operator, is given
below:
Note that the error number will not be presented on the screen.
-7001, “Entry contains illegal characters”
-7002, “Entered value is outside the allowed range”
-7003, “Day entry is not a valid number”
-7004, “Day separator is incorrect”
-7005, “Month entry is not a valid number”
-7006, “Month separator is incorrect”
-7007, “Century entry is not a valid number”;
-7008, “Year entry is not a valid number”
-7009, “Year separator is incorrect”
-7010, “Month entry is not a valid month”
-7011, “Day entry is not a valid day”
-7012, “Hours entry is not a valid number”
-7013, “Minutes entry is not a valid number”
-7014, “Entry does not give a valid time setting”
-7016, “Bus address must be within the range 0 - 30”
-7017, “Safety voltage must be within the range 10V and 110V”
-7018, “Borderline value must be within the range 10.00 - 99.99%”
-7019, “Entry does not match previous password entry”
-7020, “New string contains illegal characters or values.”
-7024, “ WARNING Instrument near cal due date”
-7025, “ WARNING Instrument past cal due date”
-7026, “Ref frequency must be within the range 0.1 Hz and 1.3 GHz”
-7027, “Ref frequency must be within the range 0.1 Hz and 3.2 GHz”
-7029, “Ext Ref must be locked for selection to be allowed”
-9001, “No calibration necessary for this function”
-9002, “Target factor is corrupt - select defaults”
-9008, “Edge speed outside limits”
-9009, “Restricted period for this amplitude”
-9010, “Maximum limit”
-9011, “Minimum limit”
-9012, “Outside amplitude range”
-9013, “Maximum amplitude for 50
Ω
load”;
-9014, “Outside frequency range”
-9015, “Outside amplitude range for DC”;
-9016, “Outside Period range”
-9017, “Outside deviation range”
-9018, “Deviation restricted by frequency range”
-9019, “Deviation restricted by amplitude range”;
-9020, “Syntax error”
-9021, “Outside numeric range”