High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Verification by Functions
57
Note
This factor applies only when the Standards DMM input is AC-Coupled
through a large capacitance. It is based on the use of a model 4955 set to
'SQV 50
Ω
'.
6.
Amplitude: a.
Measure the RMS Output Voltage value.
b.
Record this value in the Measured Value column
of the copy of the Table.
c.
For SGN1-6 calculate and record the Calculated
Values in the copy of the table.
d.
For SGN1-6 check that the Calculated Value is at
or between the RMS Limits.
e.
For SGN7-11 check that the Measured Value is at
or between the RMS Limits.
7.
9500B:
Set Output OFF.
Table 11. Sine Verification into 50
Ω
Load
Verif.
Point
Freq.
Output
Voltage
(pk-pk)
Limits (pk-pk)
Output
Voltage
(RMS)
Limits (RMS)
Measured
Value
(RMS)
Calculated
Value
Lower Higher
Lower Higher
SGN1
1kHz 1.0000V -0.020V
0.020V 0.35355V -7.07mV 7.07mV
SGN8- SGN1
SGN2
1kHz 300.00mV -6.0mV 6.0mV 106.066mV
-2.12mV 2.12mV
SGN9- SGN2
SGN3
1kHz 100.00mV -2.0mV 2.0mV 35.3553mV
-0.71mV 0.71mV
SGN10- SGN3
SGN4
45kHz 1.0000V
-0.020V 0.020V 0.35355V -7.07mV 7.07mV
SGN8- SGN4
SGN5
45kHz 300.00mV -6.0mV 6.0mV 106.066mV
-2.12mV 2.12mV
SGN9- SGN5
SGN6
45kHz 100.00mV -2.0mV 2.0mV 35.3553mV
-0.71mV 0.71mV
SGN10- SGN6
SGN7
50kHz 3.0000V
2.955V 3.045V 1.06066V 1.04475V
1.07657V
NA
SGN8
50kHz 1.0000V
0.985V 1.015V 0.35355V 0.34825V
0.35885V
NA
SGN9
50kHz
300.00mV
295.50mV 304.50mV 106.066mV 104.475mV 107.657mV
NA
SGN10
50kHz
100.00mV
98.50mV 101.50mV 35.3553mV 34.8250mV 35.8856mV
NA
SGN11
50kHz 30.000mV
29.55mV 30.45mV 10.6066mV 10.4475mV
10.7657mV
NA