High Performance Oscilloscope Calibrator
Front Panel Base Calibration by Functions
103
Table 18. Recommended Sequence of Calibrations (cont.)
Sequence Function
5
DC/Square: Square Voltage (Symmetrical)
6 LF
Sine
7
Load Resistance Measurement
DC/Square — DC Voltage Calibration
Introduction
This section is a guide to calibrating the Model 9500B's DC/Square Function, DC
Voltage; using its front panel controls.
•
Calibration Equipment Requirements
•
Interconnections
•
Calibration Setup
•
Calibration Procedure
Calibration Equipment Requirements
•
The UUT Model 9500B with 9510 or 9530 Active Head.
•
A high resolution Standards DMM with DC Voltage accuracy of ±0.005 % or
better, from 1 mV to 200 V.
For example, a Model 1281 Digital Multimeter.
•
An adaptor to convert from BNC to 4 mm leads.
For example, a Model 4955 Calibration Adaptor.
•
Short, high-quality 4 mm leads.