9500B
Calibration Manual
8
erw167
This menu can be exited only by pressing one of the five screen keys.
TEST:
This key enters Test mode, displaying the following screen:
erw168
BASE:
Runs a selftest of the Base Unit.
HEADS:
Runs a selftest of all fitted Active Heads.
ALL:
Runs a selftest of the Base Unit and all fitted Active Heads.
FAST:
Runs the same confidence selftest as at Power On.
INTERFACE:
This key allows checks of the display and display memory, the front
panel keyboard, the (Procedure mode) memory card slots, the
tracker ball and printer interfaces.
Base, Heads, All and Fast Selftest
BASE, HEADS, ALL and FAST selftests follow the same format. By pressing one of
these four screen keys on the 'Select test' menu screen, the 9500B runs that selftest. The
first screen shows the type of test, the pathway under test and the number of tests
remaining.
For example: