9500B
Calibration Manual
126
Table 33. 6-GHz Levelled Sine Function: HF Flatness
Cal. Point
Voltage
Freq
Target 1
1.6000 V
1.2 GHz
Target 2
526.30 mV
1.2 GHz
Target 3
155.30 mV
1.2 GHz
Target 4
51.300 mV
1.2 GHz
Target 5
15.240 mV
1.2 GHz
Repeat the process using the Cal. Point voltage levels in the Targets 1 - 5 sequence for
each of these frequency points, for a total of 75 calibration steps:
Targets 06 - 10: 1.50 GHz
Targets 41 - 45: 2.55 GHz
Targets 11 - 15: 1.75 GHz
Targets 46 - 50: 2.65 GHz
Targets 16 - 20: 1.95 GHz
Targets 51 - 55: 2.80 GHz
Targets 21 - 25: 2.05 GHz
Targets 56 - 60: 2.90 GHz
Targets 26 - 30: 2.20 GHz
Targets 61 - 65: 3.00 GHz
Targets 31 - 35: 2.30 GHz
Targets 66 - 70: 3.10 GHz
Targets 36 - 40: 2.45 GHz
Targets 71 - 75: 3.20 GHz
Calibration Procedure: 6 GHz Levelled Sine Function: HF Linearity Above 3.2 GHz
(9560 only)
1.
9500B: Ensure that the 9500B is in HEAD CAL, Sine, 6.4 GHz Sine mode, LIN.
The following process takes you through all TARGETs in Table 34.
2.
9500B: Set Output ON.
3.
Power Meter: Select a range that gives an on-scale reading.
4.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude on the
measuring device. The conversion from power to pk-pk voltage is pk-pk Voltage =
√
(power)*20.
5.
Press ACCEPT CALIB.
6.
Select the next TARGET and return to step 4; repeat until no TARGETs remain.