9500B
Calibration Manual
62
Table 13. Sine Flatness Verification at 1 V p-p into 50
Ω
Load
Verif Point
Nominal
Output
Voltage
Output
Frequency
Measured
p-p
Voltage at
50 kHz
Specification
Relative to 50
kHz
User’s Total
Measurement
Uncertainty
Flatness
Acceptance
Limits
Measured
p-p Voltage
for Flatness
Check
(% of Output)
(Um)
Lower
Higher
Ref 2
1 V
50 kHz
SF13
10
MHz
±2.0
SF14
50
MHz
±2.0
SF15
100
MHz
±2.0
SF16
250
MHz
±2.0
SF17
400
MHz
±2.5
SF18
550
MHz
±2.5
SF19
600
MHz
±3.5
SF20†
725
MHz
±3.5
SF21†
1
GHz
±3.5
SF22††
1.5
GHz
±4.0
SF23††
2
GHz
±4.0
SF24††
2.5
GHz
±4.0
SF25††
3
GHz
±4.0
SF60†††
4
GHz
±4.0
SF61†††
5
GHz
±4.0
SF62†††
5.5
GHz
±4.0
SF63†††
6
GHz
±4.0
†: Additional verification points for: 9500B/1100 and 9500B/3200 (†); 9500B/3200 only ††)
and 9500B/3200 with 9560 (†††).