9500B
Calibration Manual
40
Note
Please copy the following table. Enter the measurements in the Measured
Value column on the copy.
Table 2. DC/Square DC (+) Verification into 1
Ω
Load
Verif. Point
Output Voltage
Absolute Limits (+DCV)
Measured Value
Lower Higher
+DC1
1.0000 mV
0.97475 mV
1.02525 mV
+DC2
1.9000 mV
1.87453 mV
1.92547 mV
+DC3
2.3000 mV
2.27443 mV
2.32557 mV
+DC4
5.0000 mV
4.97375 mV
5.02625 mV
+DC5
6.0000 mV
5.9735 mV
6.0265 mV
+DC6
19.000 mV
18.9703 mV
19.0297 mV
+DC7
23.000 mV
22.9693 mV
23.0307 mV
+DC8
50.000 mV
49.9625 mV
50.0375 mV
+DC9
60.000 mV
59.960 mV
60.040 mV
+DC10
190.00 mV
189.928 mV
190.073 mV
+DC11
230.00 mV
229.918 mV
230.082 mV
+DC12
500.00 mV
499.85 mV
500.15 mV
+DC13
600.00 mV
599.83 mV
600.18 mV
+DC14
1.9000 V
1.8995 V
1.9005 V
+DC15
2.3000 V
2.2994 V
2.3006 V
+DC16
5.0000 V
4.99873 V
5.00127 V
+DC17
6.0000 V
5.99848 V
6.00152 V
+DC18
19.000 V
18.99523 V
19.00477 V
+DC19
23.000 V
22.99423 V
23.00577 V
+DC20
50.000 V
49.98748 V
50.01252 V
+DC21
60.000 V
59.98498 V
60.01502 V
+DC22
190.00 V
189.9525 V
190.0475 V