9500B
Calibration Manual
44
6.
Amplitude: a.
Measure the RMS Output Voltage value.
b.
Record this value in the Measured Value column of
the copy of the Table.
c.
Check that the Measured Value is at or between the
Absolute Limits.
7.
9500B:
Set Output OFF.
8.
Press the WAVEFORM soft key. Select Square (negative) by pressing the
key on the right of the screen. Repeat steps (2) to (7), but using Table 5 for each
of the verification points.
9.
Press the WAVEFORM soft key. Select Square (symmetrical) by pressing the
key on the right of the screen. Repeat steps (2) to (7), but using Table 6 for
each of the verification points.
Note
Please copy the following table. Enter the measurements in the Measured
Value column on the copy:
Table 4. DC/Square Square (+) Verification at 1 kHz into 1 M
Ω
Load
Verif.
Point
2
Frequency
Output
Voltage
(pk-pk)
Absolute Limits
(pk-pk)
Output
Voltage
(RMS)
Absolute Limits
(RMS)
Measured
Value
(RMS)
Lower
Higher
Lower
Higher
+SQ1
1 kHz
600.00 mV 599.39 mV 600.61 mV 299.9979 mV 299.67 mV 300.28 mV
+SQ2
1 kHz
60.000 mV 59.93 mV
60.07 mV
29.99799 mV 29.96 mV
30.03 mV
+SQ3
1 kHz
6.0000 mV 5.984 mV
6.016 mV
2.999799 mV 2.992 mV
3.008 mV
+SQ4
1 kHz
6.0000 V
5.99399 V
6.00601 V
2.999799 V
2.99674 V
3.00275 V
+SQ5
1 kHz
60.000 V
59.93999 V 60.06001 V 29.99799 V
29.96750 V 30.02751 V
Note
Please copy the following table. Enter the measurements in the Measured
Value column on the copy.