High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Calibration Procedures
125
Calibration Procedure: 6 GHz Levelled Sine Function: HF Linearity up to 3.2 GHz (9560
only)
1.
9500B: Ensure that the 9500B is in HEAD CAL, Sine, 3 GHz Sine mode, LIN.
The following process takes you through all TARGETs in Table 32.
2.
9500B: Set Output ON.
3.
Power Meter: Select a range that gives an on-scale reading.
4.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude on
the measuring device. The conversion from power to pk-pk voltage is pk-pk Voltage
=
√
(power)*20.
5.
Press ACCEPT CALIB.
6.
Select the next TARGET and return to step 4; repeat until no TARGETs remain.
Table 32. 6-GHz Levelled Sine Function: HF Linearity
Cal. Point
Voltage
Frequency
Target 1
450.00 mV
1.2 GHz
Target 2
700.00 mV
1.2 GHz
Target 3
1.0000 V
1.2 GHz
Target 4
1.6000 V
1.2 GHz
Target 5
2.4000 V
1.2 GHz
Target 6
3.5000 V
1.2 GHz
Calibration Procedure: 6 GHz Levelled Sine Function: HF Flatness up to 3.2 GHZ (9560
only)
1.
9500B: Ensure that the 9500B is in HEAD CAL, Sine, 3 GHz Sine mode, FLAT.
The following process takes you through all TARGETs in Table 33. The process
sequentially calibrates all amplitudes at one frequency before repeating the same
amplitude calibration points at the next frequency step.
2.
9500B: Set Output ON.
3.
Power Meter:
Select a range that gives an on-scale reading.
4.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude on
the measuring device.
The conversion from power to pk-pk voltage is pk-pk Voltage =
√
(power)*20.
5.
Press ACCEPT CALIB.
6.
Select the next TARGET and return to step 4; repeat until no TARGETs remain.
7.
Press NEXT FREQ then select TARGET 1 and return to step 4; repeat until no
TARGETs remain.