High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Calibration Procedures
123
Table 29. Levelled Sine Function: HF Flatness
Cal. Point
Voltage
Frequency
Target 1
3.0000 V
50 kHz
Target 2
1.0072 V
50 kHz
Target 3
316.49 mV
50 kHz
Target 4
316.49 mV
50 kHz
Repeat the process using the Cal. Point voltage levels in the Targets 1 - 4 sequence for
each of these frequency points, for a total of 28 calibration steps:
Targets 05 - 08: 10 MHz
Targets 17 - 20: 550 MHz
Targets 09 - 12: 50 MHz
Targets 21 - 24: 800 MHz
Targets 13 - 16: 250 MHz
Targets 25 - 28: 1.0 GHz
Calibration Procedure: 3 GHz Levelled Sine Function: HF Linearity (9530 only)
1.
9500B: Ensure that the 9500B is in HEAD CAL, Sine, 3 GHz Sine mode, LIN.
2.
The following process takes you through all TARGETs in Table 30.
3.
9500B: Set Output ON.
4.
Power Meter: Select a range that gives an on-scale reading.
5.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude
on the measuring device. The conversion from power to pk-pk voltage is pk-pk
Voltage =
√
(power)
*
20.
6.
Press ACCEPT CALIB.
7.
Select the next TARGET and return to step 4; repeat until no TARGETs remain.
Table 30. 3-GHz Levelled Sine Function: HF Linearity
Cal. Point
Voltage
Frequency
Target 1
450.00 mV
1.2 GHz
Target 2
800.00 mV
1.2 GHz
Target 3
1.2000 V
1.2 GHz
Target 4
2.0000 V
1.2 GHz
Target 5
2.8000 V
1.2 GHz
Target 6
3.7000 V
1.2 GHz