9500B
Calibration Manual
78
'Mode Selection' Screen
The Mode Selection menu screen allows you to select any one of the following operating
modes:
PROC Procedure
mode
MANUAL Manual
mode
CONFIG Configure
mode
CALIB Calibration
mode
TEST Selftest
mode
Each mode is selected by pressing the appropriate screen softkey, after which a new
menu screen relating specifically to that mode will be displayed.
Note
For the calibration menu screen to appear, certain other conditions must be
satisfied — see Selection of Calibration Mode.
Selection of Calibration Mode
The following two conditions must be satisfied before the Calibration mode menu screen
can be accessed:
1.
The rear-panel Calibration Enable switch must be in the 'ENABLE' position.
2.
A valid password must have been entered via the front panel keypad.
Calibration Enable Switch
The Calibration Enable DIP-switch is accessible, using a small screwdriver, through a
recess on the Model 9500B's rear panel.
erw213
DISABLE
With the Calibration Enable switch in the 'DISABLE' position, any attempt to select
Calibration mode by pressing the 'CALIB' screen softkey will result in the screen prompt
"
Calibration switch not enabled!
" being displayed, and access to Calibration mode will
be denied.
ENABLE
With the Calibration Enable switch in the 'ENABLE' position, pressing the 'CALIB'
screen key will result in the 'Password Entry for Calibration' user prompt screen being
displayed: