High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Calibration Procedures
135
Table 42. Edge Function: 70 ps Edge: Linearity
Cal. Point
Voltage
Frequency
Edge
Target 1
25 mV
100 kHz
Rising
Target 2
40 mV
100 kHz
Rising
Target 3
70 mV
100 kHz
Rising
Target 4
75 mV
100 kHz
Rising
Target 5
120 mV
100 kHz
Rising
Target 6
220 mV
100 kHz
Rising
Target 7
250 mV
100 kHz
Rising
Target 8
400 mV
100 kHz
Rising
Target 9
700 mV
100 kHz
Rising
Target 10
750 mV
100 kHz
Rising
Target 11
1.2 V
100 kHz
Rising
Target 12
2.2 V
100 kHz
Rising
Calibration Procedure: 70 ps Edge: Speed
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20,
and that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select, and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 150 ps Edge, SPEED mode.
5.
Set 9500B's output ON, selecting TARGET 1 from Table 43.
6.
Select a measurement device range that gives an on-scale reading.
7.
Note the rise or fall time of the edge and enter it into the Edge Speed Field on the
9500B. Make sure allowance is made for the rise time of the oscilloscope:
Edge Rise Time =
√
((Measured Rise Time)2 - (Scope Rise Time)2).
8.
Press ACCEPT CALIB.
9.
Select the next TARGET in Table 43 by pressing NEXT TARGET and return to
step 4; repeat until no TARGETs remain.