9500B
Calibration Manual
130
Calibration Procedure: 500 ps Edge: Linearity
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20. and
that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 500 ps Edge, LIN mode.
5.
Set the 9500B's output ON.
6.
Select a measurement device range that gives an on-scale reading.
7.
Adjust the 9500B's output amplitude to give a measurement device reading equal
to TARGET 1 amplitude in Table 36.
8.
Press ACCEPT CALIB.
9.
Select the next TARGET in Table 36 and return to step 4; repeat until no
TARGETs remain.
Table 36. Edge Function: 500 ps Edge: Linearity
Cal. Point
Voltage
Frequency
Edge
Target 1
1.0000 V
100 kHz
Rising
Target 2
2.0000 V
100 kHz
Rising
Target 3
2.5000 V
100 kHz
Rising
Target 4
1.0000 V
100 kHz
Falling
Target 5
2.0000 V
100 kHz
Falling
Target 6
2.5000 V
100 kHz
Falling
Calibration Procedure: 500 ps Edge: Gain
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20.
and that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 500 ps Edge, GAIN mode.
5.
Set the 9500B's output ON.
6.
Select a measurement device range that gives an on-scale reading.
7.
Adjust the 9500B's output amplitude to give a measurement device reading equal
to TARGET 1 amplitude in Table 37.
8.
Press ACCEPT CALIB.