9500B
Calibration Manual
144
Exit from Head Calibration
When the necessary calibration operations are complete, return to the Calibration screen
that contains the Exit and Store Head Cal softkeys. Selecting Exit produces the reminder
message:
WARNING:
Cal data not stored:
Exit again to abandon data
Select Exit to abandon head cal data.
To store head cal data, select the Store Head Cal softkey, which produces a new screen
labelled:
Head Calibration: CHx
(where x is the head number)
Follow the on-screen instructions to:
•
Select the warning period before recalibration is due
•
Modify the cal due date
Below the displayed dates is a message that reads:
These heads have unsaved data
followed by a line that identifies the heads e.g. CH1, CH4. If there is no cal data to store,
or if all head cal data storage is complete, the NONE indicator appears.
Selecting the Store softkey displays the message:
Saving cal factors to head x (where x is the head number)
The save operation may take up to 30 seconds to complete. To save any remaining
unsaved head data, use the Channel Select softkey to change to the appropriate head, then
select Store. Repeat for each unsaved head until the NONE indicator is displayed.
Select Exit to step back through the menus to the required instrument setting.