FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019 

Page 

48

 of 

170

 

 

  

FLINTEC

 

  

   FLINTEC 

HBM 
TOLEDO 
SYSTEC 
SMA 
 
SARTOR 
RINSTR 
AVERY 
BASTER 
LM2 






 




Flintec continuous format. 
Commonly used by HBM, GSE, PT, Systec, 
Rinstrum. 
Commonly used by Toledo, Mettler Toledo. 
Commonly used by Systec. 
Commonly used by USA producers, Cardinal, Rice 
Lake etc. 
Commonly used by Sartorious. 
Commonly used by PT, Rinstrum, HBM, GSE. 
Commonly used by Avery E1205. 
Commonly used by Baster. 
Flintec LM2 (BX1 par.000=6) 

 

[  142 

CSUM 

:  NO 

      Checksum at continuous and BSI formats. 

  

  

     

  

     

  

      NO 

YES 


Checksum disable  
Checksum enable 

  

  

[  143 

CR 

:  YES 

      Carriage return at continuous formats. 

  

  

     

  

     

  

      NO 

YES 


Carriage return disable  
Carriage return enable 

  

  

[  144 

LFEED 

:  YES 

      Line feed at continuous formats. 

  

  

     

  

     

  

      NO 

YES 


Line feed disable  
Line feed enable  

[  145 

DELAY 

:  50 

     

Data output delay at BSI formats; delay between continuous format 
data. 

  

 

  

 

 

  

  

 

000 …. 999 milliseconds.  

[  15- 

ETHERNET 

   ETHERNET TCP/IP   

 

    Press     key or  

  key again to enter this menu. 

    Or press  

  key to go to the next sub-block.

 

[  151 

FORMAT 

:  NO 

      Data format of the Ethernet port 

  

 Page 
91, 
20, 
107 
  

     

  

     

  

      NO  

CONT        
FAST         
PRNT        
BSI            
MBHL        
MBLH 







Port is disabled 
Continuous data output 
High speed continuous data output 
Printout 
BSI format for PC, PLC interface 
Modbus TCP High-Low format 
Modbus TCP Low-High format 

 

 

 

 

 

  

  

Press  

 key to select output format different than Flintec 

continuous data format while the selection is CONT in the 
information display. Available functions are shown on the weight 

display after pressing  

 key sequentially. Press  

 key after 

selecting the data format. 
 
Refer to Appendix 5 for details data structure. 

 

  

FLINTEC

 

  

   FLINTEC 

HBM 
TOLEDO 
SYSTEC 
SMA 
 
 
SARTOR 
RINSTR 
AVERY 
BASTER 
LM2 






 




Flintec continuous format. 
Commonly used by HBM, GSE, PT, Systec, 
Rinstrum. 
Commonly used by Toledo, Mettler Toledo. 
Commonly used by Systec. 
Commonly used by USA producers, Cardinal, Rice 
Lake etc. 
Commonly used by Sartorious. 
Commonly used by PT, Rinstrum, HBM, GSE. 
Commonly used by Avery E1205. 
Commonly used by Baster. 
Flintec LM2 ( BX1 par.000=6 ) 

 

[  152 

IP 

:  250 

      IP address.  

  

Содержание FT-112

Страница 1: ...FT 112 D Panel Weighing Terminal Technical Manual...

Страница 2: ...Cell Connection Only FT 112D 20 4 6 3 RS232C Serial Port 21 4 6 4 RS485 Serial Port 22 4 6 5 RS422 Serial Port 23 4 6 6 Ethernet TCP IP 24 4 6 7 Profibus PB 25 4 6 8 Profinet 25 4 6 9 CANopen 26 4 6...

Страница 3: ...87 10 3 3 SmartAPP at filling 87 Peak Hold 88 10 4 1 Product entries 88 10 4 2 Start and Stop the peak hold 90 10 4 3 SmartAPP at peak hold 90 10 4 4 Digital inputs and outputs 90 Totalization 91 11...

Страница 4: ...139 14 5 2 ESI Configuration 139 14 5 3 EtherCAT Data Structure 139 CC Link 140 14 6 1 Data Format 140 14 6 2 CC Link Configuration 140 14 6 3 CC Link Data Structure 141 Powerlink 142 14 7 1 Data For...

Страница 5: ...o be informed if any errors occur FLINTEC GmbH cannot accept any liability for direct or indirect damages resulting from the use of this manual FLINTEC GmbH reserves the right to revise this manual an...

Страница 6: ...WHEN MAKING CHECKS TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM WARNING FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT T...

Страница 7: ...erspannungs Richtlinie Richtlinie 2014 30 EU EMV Richtlinie 2011 65 EU RoHS Richtlinie 2014 31 EU Nichtselbstt tige Waagen Directive 2014 35 EU Low Voltage Directive Directive 2014 30 EU Electromagnet...

Страница 8: ...ad cell only FT 112 A D converter type 24 bit Delta Sigma ratio metric with integral analog and digital filters Conversion rate Up to 1600 measurement values per second Input sensitivity 0 4 V e appro...

Страница 9: ...s a Remote output of your PLC over BSI or Modbus Analogue Output opional Voltage output 0 5 VDC 0 10 VDC Current output 4 20mA 0 20mA Resolution 60 000 steps Dynamic filter Programmable dynamic filter...

Страница 10: ...dface2x PC Software or by keys in programming mode Device identity customization Topology Line Bus Star or Tree topology depending on physical media Installation Switched Ethernet transmission with sh...

Страница 11: ...Star or Daisy chain topology depending on physical media Installation Switched Ethernet transmission with shielded twisted pair cables and RJ 45 connectors Isolation Galvanically isolated bus electro...

Страница 12: ...type is recommended Max number of networks Up to 239 Number of connected nodes per network Master station 1 Slave station 120 Isolation Galvanically isolated bus electronics Response speed Up to 4 ms...

Страница 13: ...splay of the FT 112 D Panel instruments is shown below The meanings of the announcement symbols on the display 1 6 digit 22 mm height big weighing display with sign 2 16 digit 8 mm height alphanumeric...

Страница 14: ...D Panel Programmable Function key This programmable key is set for your easy use in your application Page 56 35 81 84 86 90 94 Programmable Function key This programmable key is set for your easy use...

Страница 15: ...or parameter Increase or next block Decrease or previous block Enter the parameter to enter data Save the data after entry Go to the next parameter Print By pressing this key the weight data is transf...

Страница 16: ...Unlock the keys 1 Press key for more than 2 seconds LOCK PASSWORD prompt appears on the display 2 Enter the key lock password 3 Press key symbol disappears on the display 3 4 2 Passwords RELATED PARAM...

Страница 17: ...cables coming to the instrument shall be high quality and shielded Distance from load cell cables interface cables and DC power supply cables to power line cables shall be minimum 30 cm The separate c...

Страница 18: ...abrasive cleaners or solvents Wipe the instrument with a soft cloth slightly dampened with warm soapy water or with mild detergent Disposal In conformance with the European Directive 2002 96 EC Waste...

Страница 19: ...be as good as possible for scale reliability Use high quality and EMC certified DC power source in the cabinet Follow the requirements on cabling Refer to page 16 Cut the panel of the cabinet to inst...

Страница 20: ...your system use 6 wire cable between indicator and the junction box and short circuit these pins in junction box for better performance as shown in the Figure 4 3 Pin number Definition Description 1...

Страница 21: ...or maximum 0 5 volt drop between the instrument and junction box Pin No Definition Description 1 A RS485 for DLC interfacing 2 B RS485 for DLC interfacing 3 V Ground 4 NC Not connected 5 NC Not connec...

Страница 22: ...ty Even Odd or No default Start Stop bits 1 start bit and 1 stop bit Table 4 3 RS 232C Serial Interface Specifications Definition RS232C Pin number TXD 1 RXD 2 GND 3 Under the terminal Table 4 4 Pin c...

Страница 23: ...inter PC PLC remote display etc Data formats Continuous Fast Continuous Printer BSI Protocol or Modbus RTU Baud rate 1200 2400 4800 9600 Default 19200 38400 57600 bps Length 7 or 8 default bits Parity...

Страница 24: ...Baud rate 1200 2400 4800 9600 Default 19200 38400 57600 bps Length 7 or 8 default bits Parity Even Odd or No default Start Stop bits 1 start bit and 1 stop bit Address Programmable between 01 99 Max q...

Страница 25: ...rt Pin no Signal DIR Description 1 TX Out Differential Ethernet transmit data 2 TX Out Differential Ethernet transmit data 3 RX In Differential Ethernet receive data 6 RX In Differential Ethernet rece...

Страница 26: ...ive RxD TxD RS 485 level 9 Housing Cable Shield Ground 4 6 8 Profinet Figure 4 15 PLC Connection PROFINET Connector pin configuration RJ45 Pin Signal DIR Description 1 TX Out Differential Ethernet tra...

Страница 27: ...ng indicator and PC should be the same in cross connection Figure 4 17 Cross PC connection 4 6 9 CANopen CANopen connection is done with four wire as indicated below in Figure 4 18 The data line ends...

Страница 28: ...In Differential Ethernet receive data 6 RX In Differential Ethernet receive data 4 Not used Terminated 5 Not used Terminated 7 Not used Terminated 8 Not used Terminated Shield Chassis ground The HUB...

Страница 29: ...Differential Ethernet receive data 6 RX In Differential Ethernet receive data 4 Not used Terminated 5 Not used Terminated 7 Not used Terminated 8 Not used Terminated Shield Chassis ground The HUB con...

Страница 30: ...ted Shield Chassis ground The HUB connection cabling will be a direct connection as shown below Figure 4 26 HUB connection 4 6 14 CC Link IE Figure 4 27 PLC Connection CC Link IE Field Connector pin c...

Страница 31: ...ctures below The alibi memory SD card can be ordered from Flintec or his dealer Figure 4 28 Location of Alibi SD card slot Inserting or removing the Alibi SD card 1 Deenergize the instrument and wait...

Страница 32: ...D card at power on to initiate its function Remove the SD card 1 Deenergize the instrument 2 Gently press on the SD card to release the lock then carefully pull the card out 3 Insert the new SD card a...

Страница 33: ...ne the accuracy and the safety of your measuring system The instrument should not share power lines with noise generating parts such as heavy load switching relays motor control equipment inductive lo...

Страница 34: ...ON ZERO ERR error prompt in case of out of range Press key to start indication without zeroing and call service If the range is 3 the residual drift cannot be displayed Refer to Power on Zero paramete...

Страница 35: ...fter placing a weight on the empty scale if this feature is enabled Refer to parameter 232 at page 55 The weight should be heavier than the value entered to the parameter 518 for automatic taring Auto...

Страница 36: ...ntinuous dynamic weighing can be used for loads which have very high vibration like reactors with high speed mixers conveyors etc Single weighing operation 1 Place container on the scale 2 Press key t...

Страница 37: ...e the container on the scale 3 Press key for more than 2 seconds at basic weighing The MEMORY message will appear on the display 4 Press key to enter memory The last entered specific tare code appears...

Страница 38: ...switch Tilt switch is used to prevent wrong weighing in case of over tilting of platform Connect tilt switch to the Input 4 of the instrument and enable tilt switch by setting the parameter 51A DUE TO...

Страница 39: ...ound on the printout data The alibi memory recording principle is a loop memory The oldest record will be deleted after the capacity limit 99 999 was reached to overwrite the latest weighing You can q...

Страница 40: ...in parameter 813 2 Press numerical keys to enter date in the printout data and press key The date format entry should be as in the operation e g date entry should be DDMMYY if parameter 251 is set to...

Страница 41: ...ge in legal or to erase alibi memory if you don t need previous records legally This process runs if the calibration switch is at programming position 1 Go to the parameter 8 METROLOGY main block in s...

Страница 42: ...nged if the calibration dip switch is at ON position These parameters are marked with M symbol in the parameter table in this section If you will change any legally relate parameter or will perform ca...

Страница 43: ...access the calibration Sub block 621 Press this key to access the diagnostic parameters Main block 9 Exit from Programming Exit from programming after set up the scale and calibration as described be...

Страница 44: ...ter arriving a parameter you may change its function by pressing or keys If the value will be entered to the parameter press numerical keys then press key and key to enter the next parameter For examp...

Страница 45: ...e the selection is CONT in the information display Available functions are shown on the weight display after pressing key sequentially Press key after selecting the data format Refer to Appendix 5 for...

Страница 46: ...High format Press key to select output format different than Flintec continuous data format while the selection is CONT in the information display Available functions are shown on the weight display a...

Страница 47: ...ontinuous data output High speed continuous data output Printout BSI format for PC PLC interface Modbus RTU High Low format Modbus RTU Low High format Press key to select output format different than...

Страница 48: ...conds 139 ADDRESS 00 Address of the port 00 99 00 means data format without address 13A DUBLEX HALF Duplex on the interface HALF FULL Half duplex select for RS485 Full duplex 14 USB USB Press key or k...

Страница 49: ...RNET TCP IP Press key or key again to enter this menu Or press key to go to the next sub block 151 FORMAT NO Data format of the Ethernet port Page 91 20 107 NO CONT FAST PRNT BSI MBHL MBLH Port is dis...

Страница 50: ...15B TIMEOUT 5 Inactivity time out 00 means disable 01 60 sec 15C CR LF YES Carriage return and Line feed at continuous formats NO YES Disable Enable 15D DELAY 050 Data output delay at demand formats d...

Страница 51: ...YES Disable Enable 167 ID1 N D ID1 printing on printout NO DATA N D Disable Print ID1 data Print ID1 name and data 168 ID2 N D ID2 printing on printout NO DATA N D Disable Print ID2 data Print ID2 nam...

Страница 52: ...ogue output Default 00 0 means the minimum output is not drifted e g enter 1 0 to set output to 1 0 volt at 0 kg in 0 10 V range 193 MAXIMUM 00 0 The maximum of the analogue output 00 0 means the maxi...

Страница 53: ...next main block 191 FORMAT INTG Data format of the Profinet INTG FLOA Signed 32 bit integer no decimal point implied 32 bit float decimal point implied 192 IP 250 IP address of Ethernet port Default i...

Страница 54: ...e sub block or press key to enter the next main block 191 FORMAT INTG Data format of the EtherCAT INTG FLOA Signed 32 bit integer no decimal point implied 32 bit float decimal point implied 19 CC LINK...

Страница 55: ...nfiguration parameters or press key to enter the next block or press key to exit from programming 21 DSPLY ACUSTI DISPLAY AND ACOUSTIC Press key or key again to enter this menu Or press key to enter t...

Страница 56: ...ith key or via serial interface Preset tare and tare with key or via serial interface at gross 232 AUTO T NO Auto taring Page 34 NO YES Disabled Auto tare if the gross load is bigger than MIN TARE and...

Страница 57: ...d total Reprint Dynamic weighing start PEAK Displaying at basic weighing HOLD Displaying at basic weighing LAST UNIT Last measurement displaying Unit change 243 ID2 KEY ID2 The function of ID2 key NO...

Страница 58: ...key or key again to enter this menu Or press key to enter the next sub block 311 APPLICA NO SmartAPP application Page 78 NO CHEC CLAS FILL PEAK SmartAPP function is disabled Weight display and SmartAP...

Страница 59: ...e next sub block 321 START AUTO Start MANU AUTO SAUT PORT SmartAPP starts by pressing key or serial port or digital input SmartAPP starts if the load is heavier than empty range SmartAPP starts automa...

Страница 60: ...W Min Weight 332 FILTER 3 0 Dynamic filtering time X X seconds 34 KEY LOCK KEY LOCK Press key or key again to enter this menu Or press key to go to the next sub block 341 UP KEY USE Setpoint key lock...

Страница 61: ...lear key locking USE LOCK Not locked Locked 35 DIG INPUTS DIGITAL INPUTS appears if option installed Press key or key again to enter this menu Or press key to go to the next sub block 351 INPUT 1 NO I...

Страница 62: ...ut function if you need function different than basic setpoint output and active high level Available functions are shown on the weight display Press key after setting up the output function Refer to...

Страница 63: ...n name is displayed until pressing enter or any alphanumeric key ID data is displayed after announcing identification name 2 seconds 372 ID1NAME ID 1 Identification name entry of ID1 key Page 41 Maxim...

Страница 64: ...e acceptable limit For example if taring is activated at INDG selection it will be disabled while exiting from set up For approved scales set the parameters in main blocks 5 6 and 8 perform calibratio...

Страница 65: ...Minimum tare for automatic taring Taring can be done if loading is heavier than MIN TARE 519 MINWEIGT 20 Minimum weight to produce printout The printout is produced if the load is heavier than MIN WE...

Страница 66: ...DLC SETUP DIGITAL LOAD CELL SETUP Only FT 112D Press key or key again to enter this menu Or press key to go to the next sub block 532 QUANTITY 01 M Quantity of digital load cell Enter the quantity of...

Страница 67: ...t test weight 62 ADJUSTMENT ADJUSTMENTS Press key or key again to enter this menu Or press key to go to the next sub block 621 ZERO ADJUSTM M Zero adjustment 622 SPAN ADJUSTM M Span adjustment 623 S U...

Страница 68: ...gross value Search by tare value Search by Consecutive value 814 TRANSFER NO Transfer Alibi memory records Page 39 NO ALL No Transfer alibi memory record to the printer port 815 ALIBI ABOUT Transfer a...

Страница 69: ...ing 913 RS485 RS485 serial port testing 914 RS422 RS422 serial port testing 915 USB USB port testing 916 IN OUT Digital Input Output testing 917 DISPLAY Display testing 918 LC SIGNAL mV Load cell sign...

Страница 70: ...EFAULT LOADING Press key or key again to enter this menu Press key to return to beginning of the sub block or press key to go to the next main block 993 PARAMET DEF M Load parameter s default Calibrat...

Страница 71: ...TEC recommends test load between 50 to 75 of the capacity 5 Place the test weight on the scale 6 Press key to start span calibration WAIT message will shown on the display 10 seconds while span calibr...

Страница 72: ...n calibration WAIT message will shown on the display 10 seconds during span calibration Span Adjustment Under Load This parameter is being used to perform span adjustment of a scale without lifting th...

Страница 73: ...enter 4000 AVARAGE LC OUT XXXXXX Enter load cell output in mV V via numerical keys If the weighing system has more than one load cell calculate the mean value of load cells outputs mV V indicated on...

Страница 74: ...own Enter the gravity acceleration value of the calibration place Confirm with key 3 625 GRAVITY USAG 9 80255 prompt shown after pressing key Enter the gravity acceleration value of the place of the u...

Страница 75: ...splay and press key to save the changes into the memory Manually Addressing an Individual RC3D digital load cell The manual addressing of load cell is done to change any load cell or to change the ins...

Страница 76: ...at from item 5 until all DLCs have been adjusted 8 After end of adjustment the following sub block appears Manual Shift Adjustment IMPORTANT NOTE The shift adjustment must be performed before calibrat...

Страница 77: ...D1 memory The descriptions below are given for ID1 key The usage of ID2 is the same Enter Identification data via alphanumeric keys 1 Press the identification key 2 Enter the identification data up to...

Страница 78: ...emory 4 The Identification data in the memory appears on the display after pressing key Enter the new alphanumeric ID data up to 32 digits and press key 5 The limit values of the item appears SET 111...

Страница 79: ...bar graph 3 The digital inputs and outputs are set to the programmed application automatically You can control lambs flaps or valves with SmartAPP digital output signals Refer to page 93 Warning the...

Страница 80: ...s minimum and maximum limits sequentially RELATED PARAMETERS Sub blocks 31 and 32 Parameters 241 242 or 243 Classifying is used for sorting products into 6 ranges Classifying can be programmed to star...

Страница 81: ...cal keys Press the key to save it or press key to return operation without saving 5 The HIGH 1255 6 prompt will show you the current high tolerance Enter new value with numerical keys Press the key to...

Страница 82: ...ion by pressing the function key which is programmed as SmartAPP key or 3 Reset the classifying with digital input 2 or 4 Send the Stop SmartAPP command via serial interface 10 1 3 SmartAPP at Classif...

Страница 83: ...Sub blocks 31 and 32 Parameters 241 242 or 243 Checkweighing operation is used to check if the product weights are in tolerance The checkweighing can be programmed to start the operation automaticall...

Страница 84: ...e key to save it or press key to return operation without saving 5 The EMPTY 100 prompt indicates the current empty range of the platform Enter new value with numerical keys 6 Press the key to return...

Страница 85: ...at checkweighing announces if the load is in tolerances with backlight color with bar graph on the right side of the display and acoustically The backlight color changes automatically to indicate the...

Страница 86: ...weight value RELATED PARAMETERS Sub blocks 31 and 32 Parameters 241 242 or 243 Filling application is used for weighing materials into the container The scale is accepted as loaded if the gross weight...

Страница 87: ...100g Enter new value with numerical keys Press the key to save it or press key to return operation without saving 4 The current dribble value appears FINE 20g Enter new value with numerical keys Press...

Страница 88: ...o parameter 111 121 131 141 or 151 4 If you will use the scale only for Filling and you prefer to activate Filling operation automatically after loading set the parameter 321 to automatic filling oper...

Страница 89: ...s at peak hold are Entry type Start to follow Reduction ratio for Medium peak detecting Test ending reduction ratio Stop testing Alarm Unloaded scale limit Value FOLLOW MID DECR END DECR STOP ALARM EM...

Страница 90: ...EMORY prompt appears 2 Press key The last used Set memory code appears SET 001 3 Enter the new memory code by pressing the numerical keys and press key 4 The previous follow value is shown on the disp...

Страница 91: ...hich is programmed as SmartAPP key or 2 Reset the process with digital input 2 or 3 Send the Stop SmartAPP command via serial interface 10 4 3 SmartAPP at peak hold The SmartAPP announces the peak hol...

Страница 92: ...2 Load the container on to the pan 3 Press key for weighing material in net 4 Add the material into the container 5 Press key for totalizing and unload the container 6 Load the following container on...

Страница 93: ...ress key to display the total value 2 Press key second time when the total displays 3 Press key and Delete prompt appears 4 Press key to delete total and item weights or press key to exit without dele...

Страница 94: ...ghing related functions Remote IO s of PLC can be used for level control of material tank conveyor control solenoid control alarm etc Basic Weighing In Out Descriptions Related parameter Input 1 Zeroi...

Страница 95: ...for entering following setpoints 5 FT 112 D Panel returns to the operation after displaying setpoint 5 and entering its value 6 Press key to return operation at any step In Out Weighing Classifying C...

Страница 96: ...example Basic Peak The instrument follows the peak of the loading during the input signal This basic weighing feature is not related with the peak hold application In SmartAPP Not for legal usage Impo...

Страница 97: ...SP1 SP1 LOW SP1 HIGH 100 kg or 100 kg and 150 kg BAS HI Basic active high Output is high if absolute indicated weight is heavier than setpoint Default BAS LO Basic active low Output is low if absolut...

Страница 98: ...ow Example if SP1 SP1 LOW SP1 HIGH SP2 SP2 LOW SP2 HIGH 100 kg or 100 kg 150 kg 200kg or 200kg 225kg BAS HI Basic active high Output is high if indicated weight is heavier than setpoint Default BAS LO...

Страница 99: ...o low if indicated weight is heavier than setpoint high Returns the high if the weight is less than setpoint low VIN HI Window active high Output goes to high if indicated weight is between setpoint h...

Страница 100: ...tive low Output is low if absolute net weight is heavier than setpoint THR HI Threshold active high Output goes to high if absolute net weight is heavier than setpoint high Returns the low if the weig...

Страница 101: ...below Example if SP1 SP1 LOW SP1 HIGH SP2 SP2 LOW SP2 HIGH 100 kg or 100 kg 150 kg 200kg or 200kg 225kg BAS HI Basic active high Output is high if net weight is heavier than setpoint Default BAS LO Ba...

Страница 102: ...ut goes to low if net weight is heavier than setpoint high Returns the high if the weight is less than setpoint low VIN HI Window active high Output goes to high if net weight is between setpoint high...

Страница 103: ...LO Basic active low Output is low if gross weight is heavier than setpoint THR HI Threshold active high Output goes to high if gross weight is heavier than setpoint high Returns the low if the weight...

Страница 104: ...mmed freely to any other function For example if SP5 is not needed to control it can be programmed to the zero range to produce empty signal SPC2 Control mode 2 The digital outputs are activated as sh...

Страница 105: ...utput is activated at the displayed weight value is stable Run The digital output is activated during the weighing indicator is in operation The output is low in power on cycle and in setup modes In N...

Страница 106: ...ue by numerical keys and press the key 6 Repeat step 4 and 5 until entering values to other three limit value of the group 7 After entering the value of 5th setpoint value press key to check values ag...

Страница 107: ...The data format of continuous data output is Status Indicated Tare STX STA STB STC D5 D4 D3 D2 D1 D0 D5 D4 D3 D2 D1 D0 CR LF CHK The including of the status bytes STA STB and STC are Definition Table...

Страница 108: ...nd 3 Bits 2 1 and 0 Always 0 Always 1 Always 1 4 3 Increment 2 1 0 Decimal point 0 1 x 1 0 0 0 XXXX00 1 0 x 2 0 0 1 XXXXX0 1 1 x 5 0 1 0 XXXXXX 0 1 1 XXXXX X 1 0 0 XXXX XX 1 0 1 XXX XXX 1 1 0 XX XXXX...

Страница 109: ...d 0A hex S Z Centre of zero O Over cap U Under cap E Weight not currently being displayed space None of the above conditions R Range 1 First range 2 Second range 3 Third range N Mode G Gross weight T...

Страница 110: ...nly sent a space character in Net operation Otherwise there is no any character NET These three characters are only sent in Net operation Otherwise there is no any character LF Line Feed 0A hex CR Car...

Страница 111: ...ng remote IP and remote port number refer to parameter 19E and 19F SINGLE LINE You can transmit the printout data in single line format by pressing key This format is recommended to send the print dat...

Страница 112: ...includes commands to get data from weighing in the label design 3 Open label design software of the printer Design your label by using the commands in item 2 4 Save Compile Convert your design to EPL...

Страница 113: ...ar will be located in the structure if not 00 Command format A general description of the command is the following ADR COMMAND CHK CR LF Response format with weight A general description of the respon...

Страница 114: ...lue immediately ADR I ADR I STATUS SIGN WEIGHT VALUE Print Read the stable weight ADR P ADR P STATUS SIGN WEIGHT VALUE Load set points ADR Q SP No L SIGN SP VALUE ADR Q STATUS Read set points ADR R SP...

Страница 115: ...y lock unlock DL 1 0 DLA Print the label DJ DJA Read the last printed data DO DOA single line data format Checksum Calculation CHK is transmitted as two ASCII characters calculated with the Checksum f...

Страница 116: ...command C Clear the tare memory Command ADR C Response ADR C A Cleared and the scale is in gross mode Comments The response status is always Ack in weighing or force mode G Read voltage value of DC p...

Страница 117: ...esponse 01SSGI Stable Gross In Range 01SDGL Dynamic Gross Low voltage error Comments The response includes 3 status information STATUS 1 can be Stable or Dynamic STATUS 2 can be Gross or Net STATUS 3...

Страница 118: ...bit 1111 outputs are implemented to char F The parameter of output s must be programmed as FBUS for this feature Page 61 X Read weight value in increased resolution Command ADR X Response ADR X STATUS...

Страница 119: ...s the analogue output to 0mA or 4 V to indicate error on the analogue output Normal Range The analogue output will reflect the gross value to the programmed analogue output 4 20 mA 0 20 mA 0 5 V or 0...

Страница 120: ...ress Modbus RTU Low High In two word registers the data is stored to the registers in little endian format Least significant word is stored to the lowest register address and most significant word is...

Страница 121: ...gh res is active B14 0 Basic Peak is passive 1 Basic Peak is active B15 0 Hold is passive 1 Hold is active B16 B26 Not used B27 0 None 1 Decimal point is X XXXX B28 0 None 1 Decimal point is X XXX B29...

Страница 122: ...puts B0 Output 1 0 Passive 1 Active B1 Output 2 B2 Output 3 B3 Output 4 B4 Output 5 40045 R W 2 Basic weighing Filling Checkweighing Classifying Peak holding Setpoint 1 High Target Target Target Follo...

Страница 123: ...W 2 Zeroing Range 0 Disable 1 2 2 3 3 20 4 50 40135 R W 2 Auto Zero Tracking 0 Disable 1 0 3d 2 0 5d 3 1d 4 2d 5 3d 40137 R W 2 Tare 0 No 1 Multi tare 2 Tare only at gross 3 Preset tare 4 Preset tare...

Страница 124: ...Max 40177 R W 2 Tare type 0 Subtractive tare 1 Additive tare 40179 R W 2 Maximum tare Refer to par 526 page 65 40181 R W 2 Secondary unit Dec Description 0 g Gram 1 kg Kilogram 2 t Ton 3 lb Libre 4 No...

Страница 125: ...not On position Check the calibration DIP switch 40197 R W 10 Reserve 40207 R 2 Voltage of power supply The value is indicated with 0 1 VDC increment for DC variant 40209 R 2 Load cell millivolt value...

Страница 126: ...gross net or tare 4 If there is an error check the error code Zero Calibration procedure 1 Check the bit B0 of 40195 which should be 1 to start adjustment 2 Load the decimal 188 to 40185 to start Zer...

Страница 127: ...F3 Zero Calibration 01 10 00 B8 00 02 04 00 00 00 BC F8 CC Span Calibration Command with Span value 50000 01 10 00 B8 00 04 08 00 00 00 DC 00 00 C3 50 94 84 Total LC capacity Command with Total LC ca...

Страница 128: ...ock 15 Additionally Ethernet parameters can be set up by Indface2x PC software over Local Network Area or by Flintec s set up PC software Both software are available on www flintec com Note For Modbus...

Страница 129: ...eration mode LED State Indication Comment Off Not online No power Check power and cable Green On line data exchange Flashing Green On line clear Flashing Red 2 flash PROFIBUS configuration error Check...

Страница 130: ...ration for PLC programmers is shown in Figure 14 1 Figure 14 1 GSD Configuration GSD Configuration Description Input 2 words 1st Dword FT 112 D Panel Output to PLC Input Input 2 words 2nd Dword FT 112...

Страница 131: ...Module Status LED LED State Description Comment Off Not power or not initialized No power or Profinet module is in initialization state Green Normal operation Green 1 flash Initialized diagnostic even...

Страница 132: ...Profinet data consist of 2 pcs Input 2 words and 2 pcs Output 2 words GSDML configuration for PLC programmers is shown in Figure 14 3 Figure 14 2 Location of Hardware catalog Parameters Descriptions D...

Страница 133: ...cription Input 2 word_1 1st Dword FT 112 D Panel Output to PLC Input Input 2 word_2 2nd Dword FT 112 D Panel Output to PLC Input Output 2 word_1 1st Dword PLC Output to FT 112 D Panel Input Output 2 w...

Страница 134: ...en blinking On line initializing Green single flash Stopped Check hardware damages Green flickering Auto baudrate detection in progress Red CANopen configuration error Check EDS file B Error LED State...

Страница 135: ...and RxPDO 64 bit EDS configuration for PLC programmers is shown in Figure 14 4 Figure 14 4 EDS Configuration EDS Configuration Description TxPDO 1 4 words Unsigned Long FT 112 D Panel Output to PLC In...

Страница 136: ...te Description Off No IP address Green Online one or more connections established CIP Class 1 or 3 Green flashing Online no connections established Red Duplicate IP address FATAL error Red flashing On...

Страница 137: ...tures consist of 2 pcs Input 2 words and 2 pcs Output 2 words EDS configuration for PLC programmers is shown in Figure 14 5 and Figure 14 6 Figure 14 5 Configuration of module properties without EDS f...

Страница 138: ...ength Description Input 2 words 1st Dword FT 112 D Panel Output to PLC Input Input 2 words 2nd Dword FT 112 D Panel Output to PLC Input Output 2 words 1st Dword PLC Output to FT 112 D Panel Input Outp...

Страница 139: ...scription Off INIT Green OPERATIONAL Green blinking PRE OPERATIONAL Green single flash SAFE OPERATIONAL Red Fatal Event B ERR LED LED State Description Off No any error or no power Red blinking Invali...

Страница 140: ...words ESI configuration for PLC programmers is shown in Figure 14 7 Figure 14 7 Configuration of module properties for Beckhoff Input Output Definition Description DI TxPDO Map SubIndex 001 1st Dword...

Страница 141: ...Red Major fault Exception or FATAL event Red flickering CRC error temporary flickering Red flashing Station Number or Baud rate has changed since startup flashing In the case of red LED warning check...

Страница 142: ...t FT 112 D Panel Output to PLC Input RWr2 RWr3 2nd Dword Input FT 112 D Panel Output to PLC Inpu Remote Input RX RX0 RX31 Not used Remote Register RWw RWw0 RWw1 1st Dword Output PLC Output to FT 112 D...

Страница 143: ...lash NMT_CS_PRE_OPERATIONAL_1 Only asynchronous data Green double flash NMT_CS_PRE_OPERATIONAL_2 Asynchronous and synchronous data No PDO data b Green triple flash NMT_CS_READY_TO_OPERATE Ready to ope...

Страница 144: ...ist of 2 pcs Input 2 words and 2 pcs Output 2 words XDD configuration for PLC programmers is shown in Figure 14 9 Figure 14 9 Configuration of module properties with XDD file Data Length Description 1...

Страница 145: ...nters state EXCEPTION If NS RUN and AS ERR turn red this indicates a fatal event In the case of red LED warning check cabling configuration and station number Power off the instrument and reenergize t...

Страница 146: ...T 112 D Panel Output to PLC Input RWr2 RWr3 2nd Dword Input FT 112 D Panel Output to PLC Input Remote Input RX RX0 RX31 Not used Remote Register RWw RWw0 RWw1 1st Dword Output PLC Output to FT 112 D P...

Страница 147: ...2 In 1 Error codes of FT 112 D Panel Unit P Tare Centre of zero Gross Net SD Read command response Cmd Flg FT 112 D Panel Output to PLC Input 2nd Dword Bit Number 2nd Dword Description B31 B24 Digita...

Страница 148: ...1 SetPoint 1 Low Only available for Threshold and Window feature Refer to sub block 36 page 61 to see the availabilities 10110 22 SetPoint 2 Low 10111 23 SetPoint 3 Low 11000 24 SetPoint 4 Low 11001 2...

Страница 149: ...d input Description B31 B11 Not in use B10 0 No Error 1 The Calibration DIP switch is not On position Check the calibration DIP switch B9 0 No Error 1 Scale unstable Wait until scale become stable Che...

Страница 150: ...mmand is activated 00001 1 Zero 00010 2 Tare 00011 3 Clear 00100 4 Print 00101 5 Adjust zero calibration Calibration 00110 6 Adjust span calibration 1 00111 7 Total Load Cell Capacity 1 eCal Coefficie...

Страница 151: ...01111 15 B weighing Filling C weighing Classifying Peak hold SetPoint 1 Target Target Target Follow 10000 16 SetPoint 2 Coarse Low Low Mid Decr 10001 17 SetPoint 3 Fine High Low End Decr 10010 18 Set...

Страница 152: ...om first to second unit 1 From second to first unit 00000110 6 R W Key lock 1 2 0 Enable 1 Disable 00000111 7 W Dynamic operation 1 0 Dynamic reset 1 Dynamic start 00001000 8 W Totalization operation...

Страница 153: ...R W Stability Detection Range 1 2 0 0 3d 1 0 5d 2 1d 3 2d 4 3d 5 4d 6 Disable 01000110 70 R W Stability Time 1 2 Refer to par 517 page 64 01000111 71 R W Unit 1 2 0 g 1 kg 2 t 3 lb 4 No unit 5 N 6 kN...

Страница 154: ...1 kg 2 t 3 lb 4 No unit 5 N 6 kN 7 kLb 01010110 01011111 86 95 Not Used 01100000 96 R W APPLICA 1 2 Refer to par 311 page 57 0 No 1 CHEC 2 CLAS 3 FILL 4 PEAK 01100001 97 R W LIMITS 1 2 Refer to par 31...

Страница 155: ...s net or tare Zero Calibration procedure 1 Check the Bit 0 of Calibration Status it should be 1 set to start adjustment 2 Write Adjust Zero Calibration command and apply New CMD to start Zero calibrat...

Страница 156: ...Input 2nd Dword Bit Number TxPDO 1 T_UL1 Description B63 B56 Digital Outputs Output bit status Active 1 B55 B48 Digital Inputs Input bit status Active 1 B47 B44 Error Codes of FT 112 D Panel Bin Dec D...

Страница 157: ...mpty High Stop 10011 19 SetPoint 5 TareMa x Not used High Alarm 10100 20 Not used Empty Not used Empty Empty 10101 21 SetPoint 1 Low Only available for Threshold and Window feature Refer to sub block...

Страница 158: ...e B15 Hold status 0 Passive 1 Active B14 Basic Peak status 0 Passive 1 Active B13 High resolution status 0 Passive 1 Active B12 Centre of zero 0 Weight is out of zero range 1 Weight is in zero range B...

Страница 159: ...ng B8 0 No Error 1 Calibration load value entry Error Test weight is too small Increase the weight B7 0 No Error 1 Calibration Error Calibration loading is not enough Check test weight loading Check l...

Страница 160: ...ions B63 B56 Set Reset digital outputs B55 B48 Expanded Commands List Refer to Table 16 3 B47 B43 Not in use B42 B38 Command List Bin Dec Commands 00000 0 None command is activated 00001 1 Zero 00010...

Страница 161: ...Label number 01100 12 Quantity of M 01101 13 Grand Total 01110 14 Not used 01111 15 B weighing Filling C weighing Classifying Peak hold Set Point 1 Target Target Target Follow 10000 16 Set Point 2 Coa...

Страница 162: ...6 R W Key lock 1 2 0 Enable 1 Disable 00000111 7 W Dynamic operation 1 0 Dynamic reset 1 Dynamic start 00001000 8 W Totalization operation 1 0 None 1 M key 2 Totalization Print 3 Totalization Cancel 0...

Страница 163: ...5d 2 1d 3 2d 4 3d 5 4d 6 Disable 01000110 70 R W Stability Time 1 2 Refer to par 517 page 64 01000111 71 R W Unit 1 2 0 g 1 kg 2 t 3 lb 4 No unit 5 N 6 kN 7 kLb 01001000 72 R W Range 1 2 0 Single rang...

Страница 164: ...5 N 6 kN 7 kLb 01010110 01011111 86 95 Not Used 01100000 96 R W APPLICA 1 2 Refer to par 311 page 57 0 No 1 CHEC 2 CLAS 3 FILL 4 PEAK 01100001 97 R W LIMITS 1 2 Refer to par 312 page 57 0 VAL 1 TOL 2...

Страница 165: ...procedure 1 Check the Bit 0 of Calibration Status it should be 1 set to start adjustment 2 Write Adjust Zero Calibration command and apply New CMD to start Zero calibration 3 Check the Bit 1 of Calib...

Страница 166: ...FT 112 D Panel Weighing Indicator Technical Manual Rev 1 0 0 May 2019 Page 165 of 170 17 SEALING OF APPROVED SCALE Sealing of Load cell connector Alibi SD card sealing Sealing of Calibration switch...

Страница 167: ...he instrument reenergizes it after 30 seconds Check external load cell connection Check load cell connector in the instrument Change main board or second scale board E02 ADC ERROR ADC conversion error...

Страница 168: ...mainboard E41 ALIBI FAULT Alibi SD card has not been supplied from FLINTEC Install Flintec Alibi SD card E42 NEW ALIBI Alibi memory serial number error The new alibi SD card is installed Format the al...

Страница 169: ...from load cell Reenergize the indicator Check the DLC connection RS 485 Power supply Change load cell E84 SN NOT MATCH DLC yy 1 The address and S N of the load cell do not match Check the DLC and S N...

Страница 170: ...FT 112 D Panel Weighing Indicator Technical Manual Rev 1 0 0 May 2019 Page 169 of 170 www flintec com...