FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
41
of
170
6 P
ROGRAMMING AND
C
ALIBRATION
You will find the programming and calibration procedure of FT-112(D) Panel weighing indicator in this section.
The arrow on navigation keys indicates the function of the keys in programming menu. The basic meanings of
these keys are;
A set-up DIP switch is located near the load cell connector at the FT-112 (D) panel rear side as shown in the
figure below and its position should be “ON” (downward) to change the metrological related parameters
including calibration.
Figure 6.1- The location of calibration DIP switch.
Entering to the Programming and Calibration
Enter the programming menu as described below. Legally related parameters can be changed if the
calibration dip switch is at ON position. These parameters are marked with
M
symbol in the parameter table in
this section. If you will change any legally relate parameter or will perform calibration, set the calibration switch
to the ON position before entering to the programming.
Display
Operation
[123.456 kg]
Press
key until [
PASSWORD :
] prompts shown.
[PASSWORD]
Enter your 4-digit passport.
(Default is 1111)
[**** ]
Press
key.
[1
—INTERFACE ]
You entered to the programming main menu and the first main
block [1
—INTERFACE ] prompt appears.
Navigation
between
parameters
Exit from any
entry or block
or from
programming.
Go to
previous block
or parameter.
Next block.
Previous
block.
Go to
next block or
parameter.
Enter the
parameter to
edit it.
Parameters’
value
Exit with
previous value
Move to the
left digit.
Next option
Before option
Move to the
right digit.
Confirm and go
to the
next parameter.
DIP Switch
Description
1
Calibration SW
OFF: Locked
ON: Set- up
2
Reserve