FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
51
of
170
[ 177
LF AFTE
: +4
]
Line feed after printout.
:
:
+ = Forward, - = Backward
NO,1,2….9 : Line feed quantity after data.
Example: -2 means 2-line feed backward.
[ 178
FORM FE
: NO
]
Form feed.
NO
YES
:
:
Disable
Enable
[ 179
LEFT SP
: 3
]
Space from left of the label.
0…9
[ 17A
COPY
: 1
]
Copy quantity.
1…9
[ 19-
ANALOG OUT
]
ANALOGUE OUTPUT
(Only FT-112(D) AN)
Press key or
key again to enter this menu.
Press
key to return to beginning of the sub-block or press
key to enter the next main-block.
[ 191
TYPE
: 4-20
]
Analog output type
4-20
0-20
0-10
0-5
:
:
:
:
4 mA - 20 mA
0 mA - 20 mA
0 VDC- 10 VDC
0 VDC- 5 VDC
[ 192
MINIMUM
: 00.0
]
The minimum of the analogue output. Default 00.0 means the
minimum output is not drifted.
e.g. enter 1.0 to set output to 1.0 volt at 0 kg in 0 -10 V range.
[ 193
MAXIMUM
: 00.0
]
The maximum of the analogue output. 00.0 means the maximum
output is not drifted.
e.g. enter 9.0 to set output to 9.0 volt at Max load in 0 -10 V range.
[ 194
SOURCE
:
GRO
S
]
Source of the analogue output
GROS
INDI
:
:
Gross weight
Indicated weight
[ 195
ZERO ADJ :
]
Displays the count value of unloaded scale.
Increase or decrease by pressing
or
; or enter the value
by pressing numerical keys to change the unloaded scale output.
[ 196
SPAN ADJ :
]
Displays the count value of the full loaded scale.
Increase or decrease by pressing
or
; or enter the value
by pressing numerical keys to change the analogue output gain.
[ 197
AUTO ADJ : NO
]
Set analogue output to calibration range
NO
YES
:
:
No
Automatic adjustment between min and
max limits if changed.