FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
124
of
170
& Errors
B1
Zero calibration in process …
B2
Span calibration in process …
B3
Calibration Timeout
- Restart calibration
B4
ADC Error
- Re-energize the instrument
- If shown again, change the board.
B5
Instrument cannot be calibrating
- Check load cell cable
- Re-energize the instrument
B6
Instrument cannot be calibrating
- Load cell signal is very low or too high
B7
Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections
B8
Calibration load value entry Error
- Test weight is too small. Increase the
weight
B9
Scale unstable
- Wait until scale become stable
- Check grounding wiring
B10
The Calibration switch is not ‘On’ position.
- Check the calibration DIP switch.
40197
R/W 10
Reserve
40207
R
2
Voltage of
power supply
The value is indicated with 0.1 VDC increment for
DC variant.
40209
R
2
Load cell millivolt value
(only FT-112(D) Panel)
Millivolt value of active scale is indicated with 0.01
mV increment. For example: 2.34 mV is indicated
as integer 234 value.
40211
R/W 2
Load parameter’s
defaults
Write 0x6BB6 value to load parameter’s defaults.
40213
R/W 2
Load factory defaults
Write 0x7CC7 value to load factory defaults.
Applications
40255
R/W 2
Application
Refer to parameter [311] page 57
0
No
1
CHEC
2
CLAS
3
FILL
4
PEAK
40257
R/W 2
Limits
Refer to parameter [312] page 57
0
VAL
1
TOL
2
%
40259
R/W 2
Information display
Refer to parameter [313] page 57
0
NO
1
ID1T
2
ID2T
40261
R/W 2
SmartAPP displaying
Refer to parameter [314] page 58
0
NO
1
BAR
2
COLO
3
ALL
40263
R/W 2
Display color change
Refer to parameter [316] page 58
0
RAAY
1
YAAR
2
RBAY