FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
167
of
170
E25 FBUS SETUP E
Fieldbus set up error.
- Press enter key and re-enter the
fieldbus setup.
- Change E
2
ROM (U13).
E26 SETPOINT ERR
Setpoint (limit values) checksum
error.
- Press enter key and re-enter the
setpoint (limit) values.
- Change E
2
ROM (U13).
E28 CLOCK ERROR
Clock error.
- Change CR2032 battery.
- Change the main board.
E29 ID NAME
ERROR
ID header checksum error.
- Change main board.
- Change E
2
ROM (U13).
E32 LABEL ERROR
Checksum error of EPL printout
format.
- Press enter key and re-load the EPL
code.
- Change E
2
ROM (U13).
E34 NOT LOADED
ADC output is not changed for
the span calibration.
- Recalibrate the scale.
- Change mainboard.
E35 LC CONNECTIO
The load cell output is decreased
after loading.
- Check load cell connections.
- Check test weight loading.
- Recalibrate the scale.
- Change mainboard.
E36 ADD LOAD
The load is not enough for span
calibration.
E37 UNSTABLE
The load is not stable at
calibration.
- Wait until scale become stable.
- Check grounding wiring.
- Recalibrate the scale.
- Change mainboard.
E40 NO ALIBI SD
Alibi memory SD card is not
installed.
- Disable Alibi memory if not required.
- Check Alibi SD card.
- Change mainboard.
E41 ALIBI FAULT
Alibi SD card has not been
supplied from FLINTEC.
- Install Flintec Alibi SD card.
E42 NEW ALIBI
Alibi memory serial number error.
The new alibi SD card is installed.
- Format the alibi memory SD
card. Refer to parameter 816.
E43 ALIBI ERROR
Alibi memory could not be
initialized.
- Check alibi memory SD card
- Change main board.
E44 ALIBI CSUM E
Alibi CSUM error.
- Check alibi memory records.
- Change Alibi memory SD card.
- Change main board.
E47 ALIBI CSUM E
Alibi information CSUM error.
E48 ALIBI SD ERR
Wrong SD card at alibi memory.
- Order Alibi SD card.
E50 DLC CARD FAI
The DLC Board is broken or not
installed.
- Re-energize indicator.
- Change the DLC Board.
E61 FLASH ERROR
E2PROM Error.
- Change main board.
E70 MB SELECT ER
Modbus selection error
- Check data format of other interfaces.
- Other interfaces should not be
Modbus.
E71 TARING ERROR
Tare range error at filling
- Check scale stability.
- Check tare min and max values.
E72 TARGET ERROR
Target value error at
checkweighing.
- Check Target value.
E73 TOTAL ERROR
Totalization CSUM error.
- Restart totalization after deleting
before total.
- Change E
2
PROM (U13).
- Change main board.
E74 TOTAL PRT ER
Totalization data cannot be
recorded to the alibi memory.
- Check alibi memory SD card.
- Change main board.
E75 MODBUS FAULT
The Modbus SD card is not
installed in SD2 card slot.
- Order Modbus SD card.
- Install Flintec Modbus SD card.