FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
168
of
170
E76 QTY OVER 99
Up to 99 items can be totalized.
- Press enter key to finalize totalization
after printing.
- Press MRC key to delete totalization.
E78 NO PT RECORD
Preset tare memory error.
- PT is not entered or PT checksum
error.
- Enter PT.
- Call service
E80 VERIFY SCALE
Reverification the scale.
- Reverify the scale after checking the
scale hardware, load cells,
performance etc.
E81 CANNOT ADDR
DLC could not addressed.
- Check the DLC connection (RS-485 &
Power supply) hardware.
- Check the DLC and S/N.
E82 SHIFT ADJUST
Shift adjustment is not available
due to load cell coefficients are
out of limits.
- Check addressing is done correctly.
- Check test weight loading on
the correct DLC.
- Check the load cell installation and
scale installation.
E83 DLC COUNT ER
No regular response from load
cell
- Reenergize the indicator.
- Check the DLC connection (RS-485 &
Power supply).
- Change load cell.
E84 SN NOT MATCH
DLC yy
(1)
The address and S/N of the load
cell do not match.
- Check the DLC and S/N.
- Reenergize the indicator.
- Readdress the DLC.
E85 DLC TIMEOUT
DLC yy
(1)
Communication time out
- Check the DLC connection (RS-485 &
Power supply) hardware.
- Check the DLC and S/N.
E86 DLC COM ERR
DLC yy
(1)
Status error of load cell
- Change load cell.
E87 DLC UNDER
DLC yy
(1)
The DLC is under
- Check mechanical installation and
DLC.
E88 DLC OVER
DLC yy
(1)
The DLC is over
- Check mechanical installation and
DLC.
E89 DLC CHKSUM
DLC yy
(1)
Checksum error
- Check termination resistors.
- Check the DLC connection (RS-485 &
Power supply) hardware.
E90 DLC PWR ERR
Power supply of DLCs could not
detected.
- Check the main board.
E91 DLC SYSTEM
Internal communication error
between the DLC board and the
indicator.
- Change the DLC board.
- Change the main board.
E92 DLC UNMATCH
Capacity of the load cell is
different.
- Check capacity of the DLC
1) yy = Number/Address of the digital load cell.