FT-112(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
57
of
170
[ 26-
PASSWORDS
]
PASSWORD ENTRIES
Press key or
key again to enter this menu.
Press
key to return to beginning of the sub-block or press
key to enter the next main-block.
[ 261
KEY
LOCK
: 11
]
Key lock password.
To lock keys and to erase grand total. NEW
Page 15
[ 262
USER
: 1111
] Set up password.
Page 15
NEW - CONFIRM
[ 263
SERVICE : 1111
] Set up password.
Page 15
NEW - CONFIRM
[ 3-- APPLICATION ] Application Block
[ 3--
APPLICATION
]
APPLICATION RELATED PARAMETERS MAIN BLOCK
Press
key sequentially to access this main block,
or press key or
key to enter configuration parameters,
or press
key to enter the next block,
or press
key to exit from programming.
[ 31-
SMARTAPP
]
SMARTAPP OPERATION SET UP
Press key or
key again to enter this menu.
Or press
key to enter the next sub-block.
[ 311
APPLICA
: NO
]
SmartAPP application.
Page 78
NO
CHEC
CLAS
FILL
PEAK
:
:
:
:
:
SmartAPP function is disabled.
Weight display and SmartAPP will
operate as check weighing,
Weight display and SmartAPP will
operate as classifying
Weight display and SmartAPP will
operate as filling,
Weight display and SmartAPP will
operate as peak holding,
Page 82
Page 79
Page 85
Page 88
[ 312
LIMITS
: TOL
]
SmartAPP limits entry type.
VAL
TOL
%
:
:
:
The weight value entry
Absolute deviations from Target
Percent deviation (relative
deviation) from target value.
[ 313
INFODIS
: NO
]
Information display at SmartAPP operation
NO
ID1T
ID2T
:
:
:
No any application
message
ID1 data and Target
ID2 data and Target
[ 314
DISPLAY
: ALL
]
SmartAPP displaying.
Page 78
NO
BAR
COLO
ALL
:
:
:
:
SmartAPP displaying is disabled.
Only bar graph operates.
Only display color warns the operator.
Bar graph and display color functions are activated.