Troubleshooting Guide
Error Messages
9-4
| FLIGHT 60 Service Manual
Displayed error
Potential Cause
Suggested Action
HIGH MV INS
Air leakage
Check internal hoses connections
HIGH MV EXH
Calibration problem
Hardware problem
Check internal hoses connections
Perform Volume factor calibration
Replace Main Board
Replace manifold
HIGH / LOW P BASE
Solenoid malfunction
Check solenoid
Perform Circuit Test
Replace solenoid
PCV NOT REACHED
Calibration problem
Hardware problem
Perform Volume Factor calibration
Replace manifold
LOW MV EXH
Air leak
Hardware problem
Check internal hoses connections
Replace Main Board
MOTOR FAULT
Motor control
Motor
Replace Power Board
Replace manifold
PRESS SENSOR
Patient pressure sensor higher
than outlet
Perform Pressure sensors calibration
Replace Main Board
POWER FAULT
Check log file to identify failed
power:
3.3V POWER FAULT
Replace Main Board
MAIN BOARD 10V POWER FAULT
Check cables
Replace Power Board
Replace Main Board
POWER BOARD 10V POWER
FAULT
Check cables
Replace Power Board
5V POWER FAULT
Replace Main Board
MEMORY FAULT
Parameters mismatch
Perform OVP
Replace Main Board
MAIN BAT CHARGER
Charger malfunction
Replace battery
Replace Power Board
MAIN BAT V HIGH / LOW
Out of range
Replace battery
Replace Power Board