Safety Instructions
General Warnings
2-2
| FLIGHT 60 Service Manual
Ensure that the oxygen source is not empty before and during the use of the
optional Air/Oxygen Entrainment Mixer or Oxygen Blending Bag Kit.
As Li-Ion batteries are charged and discharged over time, their ability to hold a
charge is decreased with use. This can shorten the amount of time the
ventilator can function while on battery power.
The batteries should be replaced when the batteries no longer meet the needs
of the user. This depends on a number of factors including settings and usage
patterns.
Charge the batteries for a minimum of three hours before powering the
ventilator from the batteries. This provides fully charged batteries.
During storage, charge the batteries for a minimum of three hours every 30
days. This provides charged batteries.
Always ensure that the green Ext. Power LED is illuminated after connecting
the FLIGHT 60 Ventilator to an external AC or DC power source. If the LED is
not illuminated, check all power connections and resolve any problems.
Always plug the FLIGHT 60 Ventilator into an AC power supply source when not
in use, to ensure best battery performance.
The flow resistance of the air inlet filter, located on the right side of the
ventilator, is likely to increase with repeated use. Ensure that the filter is
changed regularly.
Only a FLIGHT MEDICAL approved patient circuit can be used with the FLIGHT
60 Ventilator.