Removing and Reinstalling Modules
Replacing the Main Board
4-18
| FLIGHT 60 Service Manual
Replacing the Main Board
Record system S/N, manifold S/N and motor hours prior to replacing the main
board. After replacing the board need to update the SW with those numbers,
as described in chapter 8.
4.10.1 Removing the Main Board
To remove the main board:
1. For iO
2
internal mixer model - Follow the procedure for Removing the O
2
mixer
neb board (see section 4.9)
2. Remove the 6 Torx screws holding the Main Board to the front panel
3. On the bottom side of the Main Board, disconnect buzzer connectors LS1 and LS2.
4. Disconnect the back light cable from connector P5.
Do not pull the wires. Press the locker and then remove.
Figure 35 – LS1; LS2; P5 connectors
LS1
LS2
P5