Software Upgrade
Installing the SW upgrade
6-4
| FLIGHT 60 Service Manual
6
Software Upgrade
Required equipment:
USB storage device with the SW version (.pkg file)
Software upgrade will not run if the batteries are charged below
50%.
Installing the SW upgrade
1.
Turn off the ventilator, disconnect from AC.
2.
Copy the new SW version (.pkg file) to a USB storage device.
3.
Plug the USB storage device to the vent USB port.
4.
Connect the ventilator to AC.
5.
The ventilator powers up and a software update screen opens, displaying the
current version and the new version to update.
6.
Press
OK
on the touch screen.
7.
Once upgrade is done, the ventilator displays " Finished updating, please remove
DOK"
8.
Remove the USB storage device and press
OK
on the touch screen.
9.
Turn on the ventilator, press
Technical
/
and press
Show Info.
10.
Verify the new software version is displayed.
Calibration integrity check
1.
Set the ventilator to the same ventilation modes:
ACMV
VCV = 500 mL
PEEP = 5
P trig = -5.0
Rate
= 15
Ti
= 1
O
2
= 80%
Waveform = Square
2.
Connect to a patient circuit. Perform circuit test.
3.
Start ventilation.
4.
Verify the following: