Removing and Reinstalling Modules
Replacing the Manifold Assembly
Flight 60 Service Manual
|
4-23
7.
Disconnect the inlet assembly from the flex tube by pulling out the tube.
Figure 43 – Pulling out the Flex Tube
8.
Disconnect the O
2
Mixer hose from the muffler
9.
Cut the cable ties securing the muffler.
Figure 44 – Cutting the Cable Ties
10.
For iO
2
model - Disconnect the vertical silicon tube from the muffler
Figure 45 – Silicon tube