![Flight Medical Innovations 60 DL Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/flight-medical-innovations/60-dl/60-dl_service-manual_2305321062.webp)
Removing and Reinstalling Modules
Replacing the Solenoid Board Assembly
4-32
| FLIGHT 60 Service Manual
Figure 57 – Disconnecting the Check Valves
5. Remove the four Philips screws (star washers and flat washers) from the purge
board. Carefully remove the stand-offs between the purge board and the solenoid
board.
6. Carefully separate (pull out) the purge board from the connector on the solenoid
board.
4.17.2 Installing the Purge Board Assembly
To install the purge board assembly:
Follow the above procedure in reverse order.
Replacing the Solenoid Board Assembly
4.18.1 Removing the Solenoid Board Assembly
To remove the solenoid board assembly:
1. Remove the detachable battery (see section 4.3) and the internal battery (see
section 4.4).
2. Remove the ventilator cover (see section 4.5).
3. Remove the front panel (see section 4.8).
4. Remove the purge board (see section 0).
5. Disconnect connector P1 and the flat cable from the solenoid board.