Service/Technical Menu
Advanced Screen
Flight 60 Service Manual
|
5-1
5
Service/Technical Menu
To enter the Service/Technical menu:
1.
Press
Technical
/
2.
Press
ADVANCED Screen
.
3.
Enter the code 5844, and confirm by pressing
Advanced Screen
Button
Options
Description
Enable B-LEVEL
ON/OFF
Enable/Disable B-LEVEL ventilation mode
Enable O2 mixer
O2 Only
Nebulizer is disabled
O2+Neb.
Mixer + Nebulizer
OFF
Universal Circuit
ON/OFF
Flow sensor less patient circuit (flow sensor is
Integrated in the DL exhalation valve)
Purge Period
60-300
Time interval (sec.) for purge pump
operation
Additional Config
Quick Start
Enable/Disable quick start buttons
Display C & R
Enable/Disable lung mechanics values
Japanese config
Screen Intensity
Value
Increase/decrease the screen intensity
Language
Choosing the language
Enable PRVC
ON/OFF
Enable/Disable PRVC ventilation mode