Removing and Reinstalling Modules
Introduction
Flight 60 Service Manual
|
4-1
4
Removing and Reinstalling Modules
Introduction
This chapter provides detailed procedures on how to remove FLIGHT 60 Ventilator
modules and assemblies that you suspect are worn, damaged, or performing
improperly, and how to install new or repaired modules and assemblies.
Make sure you are familiar with the FLIGHT 60 Ventilator's operation before
attempting any service or maintenance.
Ventilator log files must be downloaded prior to any service attempt.
Refer to the Maintenance chapter for instructions.
Hazardous voltages are present inside the ventilator.
Disconnect
electrical power and oxygen sources before attempting any
disassembly.
Remove watch and any jewelry. Failure to do so may
result in injury to service personnel or damage to the ventilator.
Required Equipment
The following equipment is required to remove and replace FLIGHT 60 ventilator
modules:
Antistatic wrist strap
Phillips screwdriver PH1
Flat screwdriver 0.8x4
Torx screwdriver BT10x200
Cutter
Cable ties: PLT1M-M69; PLT4S-M69
To protect external parts and the touch screen from any damage while in
contact with the working surface, place the ventilator on a cushioned surface
while it is being serviced.
After servicing the ventilator modules, refer to the Test Table in the Operational
Verification Procedure (OVP) chapter and perform the necessary Operation
Verification Tests.