Calibration and Verification tests
Operational Verification Procedure (OVP)
7-20
| FLIGHT 60 Service Manual
the
EXT PWR
indicator is not lit (external power)
the battery (
BAT
) LED indicator is lit
3.
Verify that the ventilator functions normally with no further alarms.
4.
After at least 90 seconds have passed, reconnect to AC power.
5.
Verify that within 30 seconds of the connection time the ventilator indicates
switching back to AC power.
6.
Verify that the following occur:
the battery (
BAT
) LED indicator is off
the green
EXT PWR
is lit
7.2.11 Inlet Leak
To test the inlet leak:
1. Set the ventilator to STS settings.
2. Connect the ventilator to a test lung using a patient circuit.
3. Verify that the three screws securing the filter cover are tight, to insure that air
leakage is prevented.
4. Connect a vacuum meter to the air inlet.
5. Activate ventilation.
The vacuum reading should be at least -0.16 (negative 0.16) Bar at the end of
expirium.
During expiration, a decrease in the vacuum meter’s negative pressure is
normal.
If the test fails:
1. Tighten the screws locking the filter cover.
2. Replace the filter.
3. Remove the filter cover and check the O-ring.
4. Check the Inlet-to-muffler connection for a leak.