Calibration and Verification tests
Operational Verification Procedure (OVP)
Flight 60 Service Manual
|
7-21
7.2.12 Buzzer Test
To test the buzzer:
Press
Parameters
/
and choose
Screen ALARMS
.
5. Tap the Buzzer button.
6. Choose HIGH and press
Enter
/
. Verify that a High buzzer sound is heard.
7. Tap the Buzzer button.
8. Choose LOW and press
Enter
/
. Verify that a Low buzzer sound is heard.
If the test fails:
Check the buzzers cables. Make sure they are connected to the Main
Board and are intact.
Replace the Main Board (see 4.10), and perform the required calibrations
(see 7.3 Test table).
7.2.13 Oxygen Calibration Test
For non iO
2
ventilator (no internal O
2
mixer)
Connect an external Air/Oxygen Entrainment Mixer to an Oxygen source. Install the
Air/Oxygen Entrainment Mixer on the ventilator’s air inlet.
O
2
% settings are changed with the mixer dial.
For iO
2
ventilator (with internal O
2
mixer)
Connect the ventilator to an Oxygen source.
O
2
% settings are changed by the %O
2
button on the touch screen
1. Press
Technical
/
and then press
O
2
sensor
.
2. Select
ON
.
3. Set the Low
FiO
2
and the High
FiO
2
alarm to
OFF
.
4. Connect the patient circuit via a calibrated Oxygen Analyzer to the test lung.
5. Set the ventilator to STS mode. Start ventilation.