Removing and Reinstalling Modules
Replacing the Touch Screen
Flight 60 Service Manual
|
4-21
Figure 40 – Removing the J3 Flat Cable
10. Remove the Main Board.
Do not remove the silicon tubes from the main board when it is sent for repair.
4.10.2 Installing the Main Board
To install the main board:
1. Check that the main board sockets do not have any bent pins.
2. Perform the removal procedure in the reverse order.
Replacing the Touch Screen
The touch screen cannot be detached from the front panel; When the touch screen
needs replacement , the Front Subassembly (P/N V60-70010-60) is to be replaced
1. Follow the procedures for removing the main board (see section 4.10)
2. Replace the front Subassembly
To reassemble the touch screen:
Perform the above procedure in reverse order.