Removing and Reinstalling Modules
4-28
| FLIGHT 60 Service Manual
Figure 52 – Removing the Oxygen Sensor
5. Turn the oxygen sensor counterclockwise until it comes out.
4.13.2 Installing the Oxygen Sensor
To install the oxygen sensor:
Follow the above procedure in reverse order.
Replacing the O
2
leak sensor
*
(
iO
2
Internal mixer
model only
)
4.14.1 Removing the O
2
leak sensor
To remove the O
2
leak sensor:
1.
Remove the ventilator cover (see section 4.5).
2.
Disconnect the electrical cable from the O
2
leak sensor.
3.
Turn the O
2
leak sensor counterclockwise until it comes out.
4.14.2 Installing the O
2
leak sensor
To install the oxygen sensor:
Follow the above procedure in reverse order.