Calibration and Verification tests
Calibration
Flight 60 Service Manual
|
7-7
Use a Philips screwdriver to adjust the relief valve by tightening or
loosening the calibration screw located on the underside of the outlet
fitting.
Figure 72 –Adjusting the Pressure Relief Valve
10.
Press
Enter
/
to confirm the reading and to complete the calibration.
If the test fails:
Check internal tubing connections.
7.1.2 Pressure Sensors Calibration
To calibrate the pressure sensors:
1. Connect the ventilator to an external pressure meter using a reusable patient
circuit.
2. Connect the pressure meter to the test lung.
3. Use the following settings for the test lung:
C
= 10
Rp
=
5
4.
Press
Technical
/
and then press
ADVANCED Screen
.
5.
Enter the code
5844
, and confirm by pressing
6.
Press
SERVICE SCREEN
7.
Disconnect the patient circuit from the ventilator to verify there is no pressure
buildup in the circuit. Reconnect the circuit to the ventilator